EC Tests at UVa, 2015
Summary and Next step:
- Achieved sub-100 ps timing resolution with LASPD bar coupled to 55 degree light guide using optical cement.
- Reproduced IHEP preshower tile results of about 70 photoelectrons, and achieved a record 78 photoelectrons.
- Test tile response with DDK fiber connectors.
- Determine and apply position-dependent time-walk corrections.
September 25th - October 1st, 2015
- Time resolution measurements using 3-bar test with Fine-mesh PMT:
Using the EJ-200 plastic scintillators (5 cm x 5 cm x 30 cm) with each end coupled with
optical grease to a R9779 PMT for trigger. The middle scintillator bar was viewed by a R9779 PMT on the left-side
of the bar and a 2-inch fine-mesh (H6614-70) PMT on the right-side of the bar.
- The HV settings for the PMTs are as follows:
- Top right HV = -1590 V
- Top left HV = -1025 V
- Middle right HV = +1250 V
- Middle left HV = -1425 V
- Bottom right HV = -1000 V
- Bottom left HV = -1235 V
- For run 1591, the HV of the FM-PMT was increased to 1450 V.
- For the set of measurements, the bars were stacked vertically with respect to each other.
The three trigger bars were oriented parallel to each other. The vertical spacing is about
24.2 cm between the top of the top bar and the top of the bottom bar. The trigger was generated
by forming an "AND" of the left and right PMT signals of the top and bottom scintillator bars.
The top right bar's signal was delayed by 16 ns with respect to the other signals so that this PMT
would always carry the trigger time. The middle bar was not part of the trigger.
- For all measurements, the PMT signals were split from the dark box with one signal passed to a
discriminator channel set at -50 mV. The other signal was sent through 80 feet (123 ns)
of delay cable before being sent to the QDC module.
- Conclusions and Observations:
- The timing resolution of the three bar measurments was determined using the procedure discussed in the
CLAS12 FTOF report.
- For each run, the QDC pedestals were deterimined in order to subtract the offset from the QDC response. The
pedestal was isolated by placing an anti-cut on the TDC for the corresponding QDC channel. Table 15 summarizes
the pedestal and MIP response for each PMT.
Table 15: Timing resolution data using three bar method with five
Hamamatsu R9779 PMTs and one H6614-70 fine-mesh PMT
run#
|
Discriminator threshold [mV]
|
Pedestals
|
measured MIP positions
|
comments
|
1590
|
-50
|
file1
|
file2
|
Trigger: "AND" of four pmts
top and bottom bars |
1591
|
-50
|
file3
|
file4
|
Trigger: "AND" of four pmts
top and bottom bars |
- Then the time from the TDC was plotted versus the pedestal subtracted QDC response (1/sqrt(QDC - pedestal)): runs
1590 and 1591.
The top right PMT is self-timed, and hence there is no time dependence wrt the QDC signal.
- A profile histogram was created so that the distribution could be fit: runs
1590 and 1591.
The slope from the fit was used to correct for the time-walk effect.
- A cut was also placed on the pedestal subtracted QDC distributions to remove the low energy
events.
For run 1590, the following QDC cuts were chosen:
- Top right cut = 290 channels
- Top left cut = 330 channels
- Middle right cut = 300 channels
- Middle left cut = 840 channels
- Bottom right cut = 290 channels
- Bottom left cut = 280 channels
For run 1591, the following QDC cuts were chosen:
- Top right cut = 260 channels
- Top left cut = 330 channels
- Middle right cut = 1020 channels
- Middle left cut = 840 channels
- Bottom right cut = 290 channels
- Bottom left cut = 280 channels
- The initial time-walk slopes were optimized by varying one slope and holding
the other 5 fixed until the timing resolution was minimized. This procedure was repeated for each PMT.
- Using the expression T = 0.5*(tt + tb) - tm, where tt, tm, and
tb represent the time the particle passes through the top, middle, and bottom bars, respectively, the time
resolution of the bars was determined.
- The time can be expressed in terms of the times of PMT signals as T = 0.25*(ttr + ttl
+ tbr + tbl) - 0.5*(tmr + tml), where the subscript 'r' and 'l' denote the
right-side and left-side pmts, respectively, for that particular bar.
- Table 16 contains a summary of the time resolution results. The σ from the fit
represents the convolution of all six PMTs in the system. The reference counter (middle scintillator bar)
resolution is obtained by σm = sqrt(σT2
- 0.5*σr2), which is the value shown in Table 16 with σr = 58 ps from
previous measurements.
- The timing resolution of the fine-mesh PMT can be deconvoluted by using
σm2 = 0.25*(σmr2 + σml2), and
σml = sqrt(2)* σr = 82 ps, since the left-side PMT is also a R9779.
Solving the expresion for σmr (the fine-mesh PMT), yields
σmr = σFM-PMT = 104 ps (run 1590) and
σmr = σFM-PMT = 83 ps (run 1591).
- The higher gain on the FM-PMT for run 1591 helps to reduce the sensitivity to the time-walk effect, and we obtain
the same timing resolution as achieved when using all R9779 PMTs.
- Table 17 shows the time response for the Hamamastsu R9779 and H6614-70 PMTs. The TTS is about a factor of 1.75 worse for
the fine-mesh PMT.
- Using the achieved time resolution for each PMT and the light decay constant of the scintillator
(τsct = 2.1 ns), the number of photoelectrons can be estimated from σPMT2
= (σTTS2 + τsct2)/NPE, which is about 640-660 photoelectrons
for each PMT.
Table 16: Timing resolution results using three bar method with Hamamatsu R9779
PMTs and 2-inch FM-PMT
run#
|
Raw Resolution [ps]
|
Time Walk Corrected Resolution [ps]
|
Time Walk Corrected with QDC cuts Resolution [ps]
|
Time Walk Corrected with High-amp QDC cuts Resolution [ps]
|
1590
|
T Spectrum: 250
|
79
| 66
|
65
|
1591
|
T Spectrum: 213
|
71
| 58
|
57
|
Table 17: Hamamatsu R9779 and H6614-70 (2-inch FM-PMT) Time Response
PMT Type
|
Anode Pulse Rise Time [ns]
|
Electron Transit Time [ns]
|
Transit Time Spread [ns]
|
R9779
|
1.8
|
20
| 0.25
|
H6614-70
|
2.5
|
9.5
| 0.44
|
April 10th - 21st, 2015
- MIP Response and Time Resolution of FASPD bars:
Using the 5-mm thick scintillators from Eljen Technology in the shape of a trapezoid. The
smaller bar
is 232 mm long with sides 100 mm and 120 mm wide, and the larger bar is 444 mm long.
- The bars were first wrapped in aluminized mylar, and the covered in black tape.
- Optical grease was applied inside the scintillators' grooves.
- For the smaller bar, one 1.0-mm diameter, Y11 (200 ppm) fiber was used. The fiber was
cut from the 500-m spool with a length of 212 cm. After the fiber was embedded into the
scintillator groove, the fiber ends were cut off so that only 20 cm of fiber was not
embedded in the tile. The fiber ends were then polished with polishing paper from JLab.
- The fiber inside the groove has 3 turns for an approximate embedded length of 168 cm and a
total length of 208 cm, which includes the length of fiber outside the scintillator.
- For the 444-mm long bar, two 1.0-mm diameter, Y11 (200 ppm) fiber were used. The fibers were
cut from the 500-m spool with a length of 240 cm. After the fiber was embedded into the
scintillator groove, the fiber ends were cut off so that only 10 cm of fiber was not
embedded in the tile. The fiber ends were then polished with polishing paper from JLab.
- The each fiber inside the groove has 2 turns for an approximate embedded length of 220 cm and a
total length of 240 cm, which includes the length of fiber outside the scintillator. The combined
embedded length for the two fibers is 440 cm.
- After the fibers were embedded, the bars were first wrapped in aluminized mylar,
and then covered in black tape.
- Optical grease was coated onto the pmt cathode surface, and the four fiber ends were placed up
against the coated cathode. The fibers were held rigid by placing them inside of a DDK connector
ferrule.
- Conclusions and Observations:
- The new data can be compared to the SDU tile tests from January begining at
bullet #7, in particular, the results in Table 8 for runs 1525 and 1526. In that configuration, a
1-mm diameter fiber of length 111 cm was used with 2.5 turns. The number of photoelectrons, averaging
the two results together is 11.2.
- Using an attenuation length of 3.5 m for the Kuraray fibers, we expect an reduction in amplitude by about
25%. However, the number of fiber turns increased from 2.5 turns to 3.0 turns, which will partially
compensate the loss due to attenuation. The measured result is a loss of about 18% between the two measurements.
- For run 1554, the signal from the FASPD bar was sent through the 50-50 splitter to measure the timing
resolution. The procedure outlined in mid-January was followed to determine the timewalk
corrections and the timing resolution using the three bar method. However in this case, the FASPD was only
viewed by one PMT. The equations were adjusted to account for this fact. Also the FASPD's light was collected
by a XP2262, whereas the trigger bar PMTs are R9779's. The XP2262 has slightly worse timing resolution. The
achieved timing resolution after time-walk corrections is about 2.5 ns.
- The N.P.E. for run 1554 appears to be consistent with runs 1525 and 1526. However, the S.P.E. is sitting
on the tails from the pedestal and MIP in 1554. More than likely the true location is a bit higher.
- Applying cuts on the QDC spectra, did not have much affect on the timing resolution. The spectra can be
viewed from the link in Table 14.
- Run 1555 is with the 444 mm long FASPD bar. The PMT signal was sent through the 50-50 splitter, and the PMT HV
was increased to -2200 V in an attempt to help isolate the single P.E, though the SPE was not isolated.
- Run 1556 is with the 444 mm long FASPD bar, but the 50-50 splitter was bypassed and the PMT HV was decreased
back to -2100 V. The SPE is seen, but it is sitting on top of the tail of the multi-photoelectron peak. Normally,
the SPE at this gain is between channel 29 to 31, which would give 10.4 to 11.2 photoelectrons for the MIP.
Table 14: FASPD MIP Response and Timing Resolution Results
run#
|
Bar Length (cm)
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
Time resolution (ns)
|
comments
|
1552
|
23.2
|
1 fiber 3 turns
|
168
|
30.9
|
QDC Spectra: 295 |
9.5 |
---- |
XP2262 PMT, S/N 24561
HV = -2100 V |
1554
|
23.2
|
1 fiber 3 turns
|
168
|
13.1
|
QDC Spectra: 145 |
11.1 |
T Spectrum: 2.5 |
XP2262 PMT, S/N 24561
HV = -2100 V 50-50 Splitter to TDC |
1555
|
44.4
|
2 fibers 2 turns each fiber
|
440
|
----
|
QDC Spectra: 261 |
---- |
N/D |
XP2262 PMT, S/N 24561
HV = -2200 V 50-50 Splitter to TDC |
1556
|
44.4
|
2 fibers 2 turns each fiber
|
440
|
35.4
|
QDC Spectra: 324 |
9.2 |
N/A |
XP2262 PMT, S/N 24561
HV = -2100 V Direct to QDC, no splitter |
April 8th - 10th, 2015
- IHEP Tile Reproducibility Checks:
Using the IHEP Preshower hexagon tile with two 1.0-mm diameter, 1.5-m long Y11 (200 ppm)
fiber. The fiber ends were cut off so that only 16.5 cm of fiber is not
embedded in the tile. The fiber ends were polished with polishing paper from JLab.
- Optical grease was coated onto the pmt cathode surface, and the four fiber
ends were placed up against the coated cathode. The fibers were held rigid by placing them
inside of a DDK connector ferrule.
- Optical grease was also applied in the grooves of the tile.
- Two different PMTs were used with serial numbers: 24561 and 26799 to check reproducibility and
the possibility that the noise seen in S/N 24561 might affect the results that were obtained with
this PMT.
- Even though the noise level appears to have increased in PMT 24561, the number of photoelectrons using the
two PMTs are consistent at the 3% level, which is within the measurement uncertainties. So the results with
PMT 24561 should be reliable.
Table 13: Reproduction of Multi-fiber test with IHEP Tile
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
comments
|
1549
|
2.5 each fiber
|
155.4
|
26.0
|
QDC Spectra: 2419 |
93.2 |
XP2262 PMT, S/N 24561
HV = -2050 V |
1550
|
2.5 each fiber
|
155.4
|
28.4
|
QDC Spectra: 2726 |
96.0 |
XP2262 PMT, S/N 26799
HV = -1950 V |
April 8th, 2015
- Yield Analysis of Preshower Tile Measurements:
Using the measurement results from the IHEP, CNCS and Kedi tiles, an analysis of the SPE
and MIP yields was performed to study the relative yield between the two peaks.
- For this analysis, only the measurements using the XP2262 PMT with serial number: 24561
were considered.
- Since run 1526 with the IHEP tile, the yield of the single photoelectron peak has grown
relative to the MIP peak. This analysis will provide a quantitative value for the size
of the change in yield across the preshower measurements.
- The yields for the SPE and MIP peaks were determined by integrating the area under the
histogram distribution. The same bin size was chosen for all runs with the following sizes:
- SPE: 16 bins with QDC values 0 to 120
- MIP: 100 bins with QDC values 0 to 4000.
- After run 1539, the HV of the PMT was from -2100 V to -2050 V, since the MIP amplitude was close to the
saturation point of the QDC. Before lowering the HV, the relative yield of the MIP had decreased by about a
factor of 10 from the earlier IHEP tile measurements. After lowering the HV, the relative yield improved so that
the decrease was between a factor of 2 to 6 compared to the earlier IHEP measurements.
- The fact that the relative yield of the MIP improved after decreasing the HV appears to indicate that the
decrease in yield might be related to the PMT being noisy. There also seems to be a trend that the PMT's
performance is worsening with time, since the relative yield appears to be slowly dropping with time.
Table 12: SPE and MIP Yields from IHEP, Keid, and CNCS Tiles
run#
|
Tile
|
SPE Yield
|
MIP Yield
|
Relative Yield = MIP/SPE
|
QDC Integration Ranges
|
1526
|
IHEP
|
353
|
13051
|
37.0
|
SPE: 10-60 MIP: 1480-4000
|
1539
|
Kedi # 1
|
1596
|
6136
|
3.8
|
SPE: 10-60 MIP: 1480-4000
|
1537
|
Kedi # 2
|
5176
|
20637
|
4.0
|
SPE: 10-60 MIP: 1480-4000
|
1542
|
Kedi # 3
|
463
|
8387
|
18.1
|
SPE: 10-53 MIP: 1000-4000
|
1540
|
Kedi # 4
|
1383
|
14508
|
10.5
|
SPE: 10-53 MIP: 1000-4000
|
1543
|
CNCS # 1
|
3038
|
24602
|
8.1
|
SPE: 10-53 MIP: 1000-4000
|
1544
|
CNCS # 2
|
720
|
9019
|
12.5
|
SPE: 10-53 MIP: 1000-4000
|
1546
|
CNCS # 3
|
3535
|
37082
|
10.5
|
SPE: 10-53 MIP: 1000-4000
|
1547
|
CNCS # 4
|
1426
|
8624
|
6.0
|
SPE: 10-53 MIP: 1000-4000
|
1548
|
CNCS # 4
|
1297
|
9935
|
7.7
|
SPE: 10-53 MIP: 1000-4000
|
March 27th - April 8th, 2015
- 2-fiber Preshower Tile from CNCS Tests:
Using the CNCS Preshower hexagon tiles with two 1.0-mm diameter, 1.2-m long Y11 (200 ppm)
fiber. The fiber ends were cut off so that only about 19-20 cm of fiber is not
embedded on each fiber end.
- Optical grease was coated onto the surface of pmt cathodes, and the four fiber
ends were placed up against the coated cathode. The fibers were held rigid using
either a DDK ferrule.
- Optical grease was applied in the grooves of the tile.
- The 5 cm thick scintillator counters used in the 3-bar tests formed the trigger. The HVs
remained the same for these counters. For the preshower tiles, two XP2262 PMTs were used with
HVs = -2050 V and -2000 V.
- Conclusions and Observations:
- The surfaces of the CNCS tiles have some scratches and blemishes.
- The width of the groove is 1 mm.
- For CNCS tiles, the groove depth is 6 mm. The grooves are 1 mm wide for both tiles. The
diameter of the groove is 9 cm.
- The length of the groove outside the circular region for the fibers to pass outside the tile
is about 2.7 cm.
- For CNCS tile #2, two of the fibers had pulled away from the cathode's surface. However, it has
been consistently seen that the system using this PMT yields a lower number of photoelectrons.
- For run 1544, the fiber ends were trimmed and repolished for CNCS tile #2, since one of the fiber ends was
bent. Also the PMTs used for the two tiles were swapped to see if the results for CNCS tile #1 can be
reproduced with the poor performance PMT.
- The performance of CNCS tile #1 greatly improved after swapping the PMT and trimming the fiber ends. On the
other hand, tile #2's n.p.e. decreased by about 38%.
- For run 1546 with CNCS tiles #3 and #4, the PMT with S/N 23963 was swapped to the PMT with S/N 26799, since the
previous test consistently shows a lower photoelectron yield using the S/N 23963 PMT. Also tile #4 had a piece of
plastic that needed to be cut away inside the groove in the region where the circular part and the straight end
come out on the left-hand side of the tile.
- For run 1547, the PMT and tile configuration was swapped. CNCS tile #3 was paired with PMT S/N 26799, and CNCS tile #4
was paired with S/N 24561 to test reproducibility with different PMTs. There was about a 9% improved for the tile #4, which
is more consistent with tiles 1 and 2. However, tile #3 dropped by about 20 photoelectrons. It was found that for tile #3,
the fibers had slide towards the top of the cathode's surface. The fibers were repositioned to the center of the cathode, and
a new run was started.
- For run 1548, after repositioning the fibers, CNCS tile #4 gained about 4-5 photoelectrons, but it seems to be mostly
related to the SPE changing from 25 to 24 QDC channels. For CNCS tile #3, 80 photoelectrons was achieved, but this is still
about 14% low compared to a high of almost 93 photoelectrons achieved previously.
Table 11: CNCS Tile, Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
run#
|
Tile
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
comments
|
1543
|
CNCS #1
|
2.5 each fiber
|
151
|
24.1
|
QDC Spectra: 2011 |
83.4 |
XP2262 PMT, S/N 24561 HV = -2050 V, Tyvek wrapping
|
1543
|
CNCS #2
|
2.5 each fiber
|
151
|
26.6
|
QDC Spectra: 1024 |
38.4 |
XP2262 PMT, S/N 23963 HV = -2000 V, Tyvek wrapping
|
1544
|
CNCS #1
|
2.5 each fiber
|
151
|
26.2
|
QDC Spectra: 1588 |
60.6 |
XP2262 PMT, S/N 23963 HV = -2000 V, PMT swapped compared with run 1543
|
1544
|
CNCS #2
|
2.5 each fiber
|
151
|
25.4
|
QDC Spectra: 2151 |
84.7 |
XP2262 PMT, S/N 24561 HV = -2050 V, PMT swapped compared with run 1543
|
1546
|
CNCS #3
|
2.5 each fiber
|
151
|
25.0
|
QDC Spectra: 2320 |
92.8 |
XP2262 PMT, S/N 24561 HV = -2050 V |
1546
|
CNCS #4
|
2.5 each fiber
|
151
|
28.9
|
QDC Spectra: 2151 |
74.4 |
XP2262 PMT, S/N 26799 HV = -1950 V |
1547
|
CNCS #3
|
2.5 each fiber
|
151
|
29.1
|
QDC Spectra: 2109 |
72.5 |
XP2262 PMT, S/N 26799 HV = -1950 V |
1547
|
CNCS #4
|
2.5 each fiber
|
151
|
25.1
|
QDC Spectra: 2032 |
81.0 |
XP2262 PMT, S/N 24561 HV = -2050 V |
1548
|
CNCS #3
|
2.5 each fiber
|
151
|
28.4
|
QDC Spectra: 2271 |
80.1 |
XP2262 PMT, S/N 26799 HV = -1950 V |
1548
|
CNCS #4
|
2.5 each fiber
|
151
|
24.2
|
QDC Spectra: 2077 |
85.8 |
XP2262 PMT, S/N 24561 HV = -2050 V |
March 20th - 27th, 2015
- 2-fiber Preshower Tile from Kedi Tests:
Using the Kedi Preshower hexagon tiles with two 1.0-mm diameter, 1.2-m long Y11 (200 ppm)
fiber. The fiber ends were cut off so that only about 20.0 cm of fiber is not
embedded on each fiber end.
- Optical grease was coated onto the surface of pmt cathodes, and the four fiber
ends were placed up against the coated cathode. The fibers were held rigid using
either a plastic block from the machine shop or a DDK ferrule.
- Optical grease was applied in the grooves of the tile.
- The 5 cm thick scintillator counters used in the 3-bar tests formed the trigger. The HVs
remained the same for these counters. For the preshower tiles, two XP2262 PMTs were used with
HVs = -2100 V and -2025 V.
- Conclusions and Observations:
- The surfaces of the Kedi tiles have some scratches and blemishes.
- For Kedi tile #1, the groove depth varies from 6 mm to 7 mm. The groove depth of Kedi tile #2
appears to be roughly uniform at 6 mm deep. The grooves are 2 mm wide for both tiles. The
diameter of the grooves is 9 cm. Tile #3 and #4 have the same dimensions and groove sizes
compared to tile #2. Kedi tile #3 has a small chipped at one of the hexagon corners, and it
has a larger chip at the middle of one of the hexagon's sides:
image1 and
image2.
- The length of the groove outside the circular region for the fibers to pass outside the tile
is about 2.4 cm compared to 3.5 cm for the IHEP tiles.
- When embedding the fibers and wrapping the tiles in Tyvek paper, the fibers were particularly
hard to keep in the grooves while wrapping. This could be due to using fibers cut directly from
500-m spool and/or due to the grooves not being deep enough.
- As was seen previously, the ferrule provides the better connectivity to the PMT surface.
- After run 1537, it was found that some of the fiber ends had an air gap between the fiber and
the cathod surface. The PMT viewing Kedi #1 was recoated with optical grease, and the fibers
were better positioned to touch the surface of the PMT's cathode. However, the SPE was difficult
to find for run 1538 with Kedi tile #1, and it appeared to shift considerably in channel number (30 ----> 40).
- For run 1539, the DDK connector was used to hold the fibers from Kedi tile #1 and the PMT used for the
measurements on Kedi tile #2 was used instead. Using the same connector for the fibers and the same
PMT results in a consistent n.p.e. (84-87) at the 3% level.
- Even though a DDK connector was used for the fibers from tile #3, the n.p.e. is low compared to the others. It was
noted that the fiber lengths were slightly different, which caused at least one fiber to pull away from the cathode's
surface.
For run 1542, the fiber ends were repositioned to obtain a better coupling to the PMT's cathod surface and the fiber
ends were trimmed to be the same length.
- For all four Kedi tiles, the maximum difference in the n.p.e. is (84-91), with an average of 87.2 and a
spread of about ± 4%.
Table 10: Kedi Tile, Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
run#
|
Tile
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
comments
|
1537
|
Kedi #1
|
2.5 each fiber
|
151
|
29.4
|
QDC Spectra: 1469 |
50.0 |
XP2262 PMT, HV = -2025 V Tyvek wrapping
|
1537
|
Kedi #2
|
2.5 each fiber
|
151
|
33.2
|
QDC Spectra: 2801 |
84.4 |
XP2262 PMT, HV = -2100 V Tyvek wrapping
|
1538
|
Kedi #2
|
2.5 each fiber
|
151
|
33.0
|
QDC Spectra: 2847 |
86.4 |
XP2262 PMT, HV = -2100 V Reproducibility of Run 1537
|
1539
|
Kedi #1
|
2.5 each fiber
|
151
|
32.0
|
QDC Spectra: 2786 |
87.1 |
XP2262 PMT, HV = -2100 V Swapped Connectors and PMTs
|
1540
|
Kedi #3
|
2.5 each fiber
|
151
|
32.0
|
QDC Spectra: 1749 |
54.7 |
XP2262 PMT, HV = -2025 V Tyvek wrapping
|
1540
|
Kedi #4
|
2.5 each fiber
|
151
|
24.2
|
QDC Spectra: 2205 |
91.0 |
XP2262 PMT, HV = -2050 V Tyvek wrapping
|
1542
|
Kedi #3
|
2.5 each fiber
|
151
|
25.0
|
QDC Spectra: 2176 |
87.0 |
XP2262 PMT, HV = -2050 V Trimmed Fiber ends
|
February 10th - 18th, 2015
- DDK Fiber Connector Tests with IHEP Tile:
Using the IHEP Preshower hexagon tile with two 1.0-mm diameter, 1.5-m long Y11 (200 ppm)
fiber. The fiber ends were cut off so that only 14.5 cm of fiber is not
embedded in the tile. The fiber ends were polished with polishing paper from JLab.
- Optical grease was applied in the grooves of the tile.
- The 5 cm thick scintillator counters used in the 3-bar tests formed the trigger. The HVs
remained the same for these counters. For the preshower tile, a XP2262 PMT was used with a
HV = -2100 V.
- For these tests, the WLS fibers were placed in a DDK connector in one ferrule, and 4.5 cm
clear fiber pieces were placed in anothre ferrule. The two ferrules were then meshed together
inside the box component of the DDK connector. The two ferrules are held in place with clips.
- For the first measurement, only optical grease was used to couple the two sets for fibers inside the
ferrules.
- For the second measurement, optical cement was applied to the two sets for fibers inside the ferrules
to keep the fibers rigid.
- For the third measurement, a third ferrule was used to optically cement one end of four 2-m clear
fibers, which was then meshed together with the ferrule holding the WLS fibers.
- Conclusions and Observations:
- The MINERvA collaboration measured average connector transmissions of about 87.0% with a standard
deviation of 4.8%:
arXiv:1305.5119v1.
- Without optically cementing the fibers into the connectors, the fibers can move, which causes less
than ideal coupling between the WLS and clear fibers.
- The measurement with with the third ferrule measures the convolution of both the DDK connector
transmission and the attenuation of the 2-m fiber length. It is assumed that the DDK connector
transmission was 84% as with the previous test with 4.5 cm clear fibers.
- A range of values is provided for the transmission, since only once has about 78 photoelectrons been acheived.
Table 9: DDK Connector Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
Transmission NPEmax = 71-78
|
comments
|
1529
|
2.5 each fiber
|
155.4
|
31.5 ± 0.8
|
QDC Spectra: 1406 |
44.7 |
57-63% |
XP2262 PMT, HV = -2100 DDK Connector using grease
|
1530
|
2.5 each fiber
|
155.4
|
31.4 ± 0.6
|
QDC Spectra: 1876 |
59.8 |
77-84% |
XP2262 PMT, HV = -2100 DDK Connector using cement
|
1532
|
2.5 each fiber
|
155.4
|
30.3 ± 0.6
|
QDC Spectra: 1229 |
40.6 |
67.8%1 |
XP2262 PMT, HV = -2100 DDK Connector using cement
coupled to 4x 2-m clear fibers
|
1Assuming 59.8 p.e. after transmission through the DDK connector. This number should represent
the
percentage of photoelectrons after attenuation through 2-m of clear fiber.
January 26th - February 10th, 2015
- Reproduction of Multi-fiber test from September 2014:
Using the IHEP Preshower hexagon tile with two 1.0-mm diameter, 1.5-m long Y11 (200 ppm)
fiber. The fiber ends were cut off so that only 20 cm of fiber is not
embedded in the tile. The fiber ends were polished with polishing paper from JLab.
- Optical grease was coated onto the pmt cathode surface, and the four fiber
ends were placed up against the coated cathode. The fibers were held rigid using
a plastic block from the machine shop.
- Optical grease was also applied in the grooves of the tile.
- The 5 cm thick scintillator counters used in the 3-bar tests formed the trigger. The HVs
remained the same for these counters. For the preshower tile, a R9779 PMT was used, though it
was found the gain was too low even at -1700 V (97% of maximum) to isolate the single
photoelectron peak from the pedestal noise.
- Additional tests were performed either with the previously used XP2262 PMT or a R11102 PMT.
- Conclusions and Observations:
- Even with the XP2262, the light collection was poor. The tyvek paper was originally wrapped on
the preshower tile in September 2014. The paper was worn and light could easily leak from around
the edges of the paper. With a fresh piece of tyvek paper, the light collection dramatically improved.
- For run 1522, the XP2262 was swapped with a R11102 set to -1400 V (93% of maximum). The relative gain
between the two PMTs based on HV setting was about a factor of 10 in favor of the XP2262. The measured
MIP position with the two PMTs was actually only a factor of 5 higher for the XP2262. However, the light
collection efficiency from the fibers to PMTs is very sensitive to connection to the PMT surface. Hence,
it cannot be guaranteed that the light collection was the same between the two runs.
- Runs 1525 and 1526 were taken back-to-back without any changes in the system. The single p.e. peak was
not promienent in run 1525, though run 1526 does have about 2.7 times more statistics. 78 photoelectrons is
an achieved record for the preshower tile.
Table 7: Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
comments
|
1516
|
2.5 each fiber
|
155.4
|
Not Found
|
QDC Spectra: 519 |
----- |
R9779 PMT, HV = -1700 V Tyvek wrapping from 09/2014
|
1518
|
2.5 each fiber
|
155.4
|
31.4 ± 2.2
|
QDC Spectra: 1617
|
51.5
|
XP2262 PMT, HV = -2100 V Tyvek wrapping from 09/2014
|
1521
|
2.5 each fiber
|
155.4
|
27.5 ± 1.2
|
QDC Spectra: 1767
|
64.2
|
XP2262 PMT, HV = -2100 V
New tyvek paper Recentered fibers on PMT
|
1522
|
2.5 each fiber
|
155.4
|
Not Found
|
QDC Spectra: 355
|
-----
|
R11102 PMT, HV = -1400 V
|
1525
|
2.5 each fiber
|
155.4
|
36.6 ± 2.1
|
QDC Spectra: 2503
|
68.4
|
XP2262 PMT, HV = -2100 V Reproducibility of Run 1521
|
1526
|
2.5 each fiber
|
155.4
|
32.1 ± 2.1
|
QDC Spectra: 2499
|
77.8
|
XP2262 PMT, HV = -2100 V Reproducibility of Run 1525
|
- 3-mm and 5-mm SDU Tile Fiber tests:
Using the SDU square tiles with one 1.0-mm diameter, 1.5-m long Y11 (200 ppm)
fiber. The fiber ends were cut off so that only 20 cm of fiber is not
embedded in the tile. The cut fiber end was polished with polishing paper from JLab.
- Optical grease was coated onto the pmt cathode surface, and the two fiber
ends were placed up against the coated cathode. The fibers were held rigid using
a fiber connector from DDK.
- Optical grease was also applied in the grooves of the tile.
- The 5 cm thick scintillator counters used in the 3-bar tests formed the trigger. The HVs
remained the same for these counters. For the SDU tiles, a R11102 PMT was used, though it
was found the gain was too low even at -1400 V (93% of maximum) to isolate the single
photoelectron peak from the pedestal noise.
- Additional tests were performed with a previously used XP2262 PMT.
- SDU tile dimensions: 10 cm X 10 cm X 0.3 cm or 10 cm X 10 cm X 0.5 cm (length X width X thickness),
the groove diameter is 8 cm with 4.1 cm long straight grooves for the fiber to pass outside the tile.
- The 3-mm tile has a 2-mm deep groove, and the 5-mm tile has a 3-mm deep groove.
- These tests were conducted in parallel with the preshower tile tests with the SDU tile lying
on top of the preshower tile.
- Conclusions and Observations:
- For run 1521, the "MIP" was isolated by cutting on the preshower tile MIP, which helped to greatly
suppress the pedestal of the SDU tile. However, it is likey that this peak is a combination of MIP and
the single p.e. peak.
- For run 1522, the R11102 was swapped with the preshower XP2262 set to -2100 V. The relative gain
between the two PMTs based on HV setting was about a factor of 10 in favor of the XP2262. The peak at
channel 30 is probably the single p.e. peak, and the small peak near channel 60 is probably the MIP
location, though there is very little signal above the single p.e. peak.
- Data for the 5-mm tile with 2.5 fiber turns were taken at two HV settings: -1900 V and -2025 V.
Both runs show a consistent result of about 11-11.5 photoelectrons.
- Data for the 5-mm tile with 1.5 fiber turns appears to indicate about 7 photoelectrons.
Table 8: SDU Tile, Fiber test results (one Y11, 1mm dia, 1.177m length)
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
comments
|
1521
|
1.5
|
45.9
|
Not Found
|
QDC Spectra: 8.26
|
-----
|
R11102 PMT, HV = -1400 V 3 mm tile
|
1522
|
1.5
|
45.9
|
30.0 ± 0.5
|
QDC Spectra: 59.7
|
2.0
|
XP2262 PMT, HV = -2100 V 3 mm tile
|
1525
|
2.5
|
71.0
|
14.7 ± 0.6
|
QDC Spectra: 167.3
|
11.4
|
XP2262 PMT, HV = -1900 V 5 mm tile
|
1526
|
2.5
|
71.0
|
33.5 ± 0.6
|
QDC Spectra: 365.2
|
10.9
|
XP2262 PMT, HV = -2025 V 5 mm tile
|
1528
|
1.5
|
45.9
|
31.0 ± 0.4
|
QDC Spectra: 227.2
|
7.3
|
XP2262 PMT, HV = -2025 V 5 mm tile
|
January 15th - 26th, 2015
- Time resolution measurements using 3-bar tests Part 3 with LASPD bar and 0 degree
and 55 degree light guides:
Using the EJ-200 plastic scintillators (5 cm x 5 cm x 30 cm) with each end coupled with
optical grease to a R9779 PMT for trigger. The middle bar is a 2 cm (thick) x 8.3 cm to 14.0 cm (wide) x 57.0 cm long bar.
The narrow bar's end was wrapped so that only the 5 cm center section was exposed to couple to the the PMT. The wide end of the
bar was attached to a light guide in the following configurations:
* 0 degree light guide with optical grease (2 cm x 14 cm to 3.9 cm diameter, 5.05 cm long with 3.9 cm diameter x 1.27 cm disc)
* 0 degree light guide with optical cement (2 cm x 14 cm to 3.9 cm diameter, 5.05 cm long with 3.9 cm diameter x 1.27 cm disc)
* 55 degree light guide with optical cement (2 cm x 14 cm to 3.9 cm diameter, with 3.9 cm diameter x 1.27 cm disc at 55 deg angle)
The same HV settings from the previous tests were used.
- The configuration of the bars was the same as the previous tests; however, the LASPD bar was approximately
centered wrt the trigger bars.
- For all measurements, the PMT signals were split from the dark box with one signal passed to a
discriminator channel set at -50 mV. The other signal was sent through 80 feet (123 ns)
of delay cable before being sent to the QDC module.
- Conclusions and Observations:
- For each run, the QDC pedestals were deterimined in order to subtract the offset from the QDC response. The
pedestal was isolated by placing an anti-cut on the TDC for the corresponding QDC channel. Table 5 summarizes
the pedestal and MIP response for each PMT. For the trigger bars, dedicated pedestal runs were taken with the aid
of a pulse generator to monitor the pedestals' locations.
Table 5: LASPD bar with Light Guides and Hamamatsu R9779
PMTs, Timing resolution data using three bar method
run#
|
Discriminator threshold [mV]
|
Pedestals
|
measured MIP positions
|
comments
|
1501
|
-50
|
ped spectra
|
MIP spectra
|
Trigger: "AND" of four pmts
top and bottom bars 0 degree light guide optical grease coupling |
1502
|
-50
|
ped spectra
|
MIP spectra
|
Trigger: "AND" of four pmts
top and bottom bars 0 degree light guide optical grease coupling |
1507
|
-50
|
ped spectra
|
MIP spectra
|
Trigger: "AND" of four pmts
top and bottom bars 55 degree light guide optical cement coupling |
1511
|
-50
|
ped spectra
|
MIP spectra
|
Trigger: "AND" of four pmts
top and bottom bars 0 degree light guide optical cement coupling |
- Then the time from the TDC was plotted versus the pedestal subtracted QDC response (1/sqrt(QDC - pedestal)):
runs 1501,
1502, 1507
and 1511. The top right PMT is self-timed,
and hence there is no time dependence wrt the QDC signal.
- A cut was also placed on the pedestal subtracted QDC distributions to remove the low energy
events.
For run 1501, the following QDC cuts were chosen:
- Top right cut = 330 channels
- Top left cut = 440 channels
- Middle right cut = 240 channels
- Middle left cut = 300 channels
- Bottom right cut = 370 channels
- Bottom left cut = 350 channels
For run 1502, the following QDC cuts were chosen:
- Top right cut = 340 channels
- Top left cut = 470 channels
- Middle right cut = 240 channels
- Middle left cut = 260 channels
- Bottom right cut = 390 channels
- Bottom left cut = 370 channels
For run 1507, the following QDC cuts were chosen:
- Top right cut = 310 channels
- Top left cut = 440 channels
- Middle right cut = 300 channels
- Middle left cut = 300 channels
- Bottom right cut = 340 channels
- Bottom left cut = 330 channels
For run 1511, the following QDC cuts were chosen:
- Top right cut = 300 channels
- Top left cut = 420 channels
- Middle right cut = 240 channels
- Middle left cut = 320 channels
- Bottom right cut = 340 channels
- Bottom left cut = 330 channels
- A profile histogram was created so that the distribution could be fit: runs
1501,
1502,
1507 and
1511. The slope from the fit was used to correct
for the time-walk effect.
- The initial time-walk slopes were then optimized by varying one slope and holding the other 5 fixed until the
timing resolution was minimized. This procedure was repeated for each PMT.
- Using the expression T = 0.5*(tt + tb) - tm, where tt,
tm, and tb represent the time the particle passes through the top, middle, and
bottom bars, respectively, the time resolution of the bars was determined.
- The time can be expressed in terms of the times of PMT signals as T = 0.25*(ttr + ttl
+ tbr + tbl) - 0.5*(tmr + tml), where the subscript 'r' and 'l'
denote the right-side and left-side pmts, respectively, for that particular bar.
- The timing distributions for data are shown in four panels in Table 6:
- the top-left panel is without any corrections
- the top-right panel includes the global timewalk correction for each PMT applied
- the bottom-left panel include the global timewalk corrections with low-amplitude QDC cuts.
- the bottom-right panel include the global timewalk corrections with low- and high-amplitude QDC cuts.
- The timewalk corrections improve the resolution by a factor of 2-3, and an additional 15-20% is achieved with the
QDC cuts. At high QDC amplitudes the statistics are poor, and the timewalk corrections extrapolate into this
region, which can cause a tail in the timing distribution. Before optimizing the time-walk coefficients, placing
high-amplitude QDC cuts could improve the timing resolution by 30-40 ps. After optimizing the coefficients, the
high-amplitude QDC cuts provide only a slight improvement of 2-5 ps.
- Table 6 contains a summary of the time resolution results for each run. The σ from the fit
represents the convolution of all six PMTs in the system. The counter (scintillator bar)
resolution is obtained from the expression: σm = sqrt(σT2
- 0.5*σr2), which is the value shown in
Table 6, where σr = 58 ps.
Table 6: LASPD bar with Light Guides and Hamamatsu R9779
PMTs, timing resolution results using three bar method
run#
|
Raw Resolution [ps]
|
Time Walk Corrected Resolution [ps]
|
Time Walk Corrected with QDC cuts Resolution [ps]
|
Time Walk Corrected with High-amp QDC cuts Resolution [ps]
|
1501
|
T Spectrum: 304
|
139
| 118
|
115
|
1502
|
T Spectrum: 323
|
141
|
121
|
119
|
1507
|
T Spectrum: 288
|
115
| 98
|
96
|
1511
|
T Spectrum: 283
|
128
|
106
|
104
|
January 9th - 15th, 2015
- Time resolution measurements using 3-bar tests Part 3 with LASPD bar:
Using the EJ-200 plastic scintillators (5 cm x 5 cm x 30 cm) with each end coupled with
optical grease to a R9779 PMT for trigger. The middle
scintillator bar was replaced with a 2 cm (thick) x 8.3 cm to 14.0 cm (wide) x 57.0 cm long bar. The
bar's ends were wrapped so that only the 5 cm center section was exposed to couple to the the PMTs.
The HVs for the six PMTs were approximately gain matched.
- The HV settings for the PMTs are as follows:
- Top right HV = -1590 V
- Top left HV = -1025 V
- Middle right HV = -1360 V
- Middle left HV = -1425 V
- Bottom right HV = -1000 V
- Bottom left HV = -1235 V
- The area of the reference bars (5 cm x 5 cm x 30 cm) is 150 cm2, and the area of LASPD bar
is 635.55 cm2. The area of the LASPD bar is about 4.25 times larger than the trigger bars.
- For the set of measurements, the bars were stacked vertically with respect to each other.
The three bars were oriented parallel to each other. The vertical spacing is about
24.2 cm between the top of the top bar and the top of the bottom bar. The LASPD bar is resting on a small
tape spool (~ 2 cm high) to raise the bar so that it is centered vertically with respect to the R9779 tubes.
The trigger was generated by forming an "AND" of the left and right PMT signals of the top and bottom
scintillator bars. The top right bar's signal was delayed by 16 ns with respect to the other signals so
that this PMT would always carry the trigger time. The middle bar was not part of the trigger.
- Since the LASPD bar is approximately twice the length of the trigger bars, two measurements will be done
where the LASPD bar is moved relative to the trigger bars. During the first measurement, the LASPD bar's
right end is located at the same position as the trigger bars' right ends. For the second measurement, the
left ends of the bars will be matched.
- For all measurements, the PMT signals were split from the dark box with one signal passed to a
discriminator channel set at -50 mV. The other signal was sent through 80 feet (123 ns)
of delay cable before being sent to the QDC module.
- Conclusions and Observations:
- For each run, the QDC pedestals were deterimined in order to subtract the offset from the QDC response. The
pedestal was isolated by placing an anti-cut on the TDC for the corresponding QDC channel. Table 3 summarizes
the pedestal and MIP response for each PMT.
- For run 1494, the first run, it was found that the QDC and TDC information was mismatched due to a DAQ error.
After run 1494, the VME crate was rebooted, and after run 1495, the CODA crl file was modified to look for and
clear any extra words in either module after an event has been readout.
Table 3: LASPD bar Timing resolution data using three bar method with Hamamatsu R9779
PMTs
run#
|
Discriminator threshold [mV]
|
Pedestals
|
measured MIP positions
|
comments
|
1494
|
-50
|
file1
|
file2
|
Trigger: "AND" of four pmts
top and bottom bars QDC and TDC information out of synchronization |
1496
|
-50
|
file3
|
file4
|
Trigger: "AND" of four pmts
top and bottom bars Right-side of bars aligned |
1498
|
-50
|
file5
|
file6
|
Trigger: "AND" of four pmts
top and bottom bars Left-side of bars aligned |
- Then the time from the TDC was plotted versus the pedestal subtracted QDC response (1/sqrt(QDC - pedestal)):
runs 1496 and
1498. The top right PMT is self-timed,
and hence there is no time dependence wrt the QDC signal.
- A cut was also placed on the pedestal subtracted QDC distributions to remove the low energy
events.
For run 1496, the following QDC cuts were chosen:
- Top right cut = 330 channels
- Top left cut = 350 channels
- Middle right cut = 320 channels
- Middle left cut = 240 channels
- Bottom right cut = 370 channels
- Bottom left cut = 340 channels
For run 1498, the MIP amplitudes were higher for three of the PMTS, and the following QDC cuts were chosen:
- Top right cut = 320 channels
- Top left cut = 430 channels
- Middle right cut = 450 channels
- Middle left cut = 400 channels
- Bottom right cut = 350 channels
- Bottom left cut = 350 channels
- A profile histogram was created so that the distribution could be fit: runs
1496 and
1498. The slope from the fit was used to correct
for the time-walk effect. For run 1496 and the middle left PMT, a third order polynomial was used instead, since there
was a transition around 0.048. Using this correction, vastly improved the timing resolution.
- Using the expression T = 0.5*(tt + tb) - tm, where tt,
tm, and tb represent the time the particle passes through the top, middle, and
bottom bars, respectively, the time resolution of the bars was determined.
- The time can be expressed in terms of the times of PMT signals as T = 0.25*(ttr + ttl
+ tbr + tbl) - 0.5*(tmr + tml), where the subscript 'r' and 'l'
denote the right-side and left-side pmts, respectively, for that particular bar.
- Added on January 27: The initial time-walk slopes were then optimized by varying one slope and holding
the other 5 fixed until the timing resolution was minimized. This procedure was repeated for each PMT.
- Updated on January 27: The timing resolution for the two runs (1496 and 1498) are quite different, and this
result is reflected in the significant increase of the MIP values for top-left, middle-left and middle-right PMTs.
A significant increase in light collection efficiency is directly related to improvement in the timing resolution.
- The timing distributions for runs 1496 and 1498 are shown in four panels in Table 4:
- the top-left panel is without any corrections
- the top-right panel includes the global timewalk correction for each PMT applied
- the bottom-left panel include the global timewalk corrections with low-amplitude QDC cuts.
- the bottom-right panel include the global timewalk corrections with low- and high-amplitude QDC cuts.
- The timewalk corrections improve the resolution by a factor of 2-3, and an additional 15-20% is achieved with the
QDC cuts. At high QDC amplitudes the statistics are poor, and the timewalk corrections extrapolate into this
region, which can cause a tail in the timing distribution. Before optimizing the time-walk coefficients, placing
high-amplitude QDC cuts could improve the timing resolution by 30-40 ps. After optimizing the coefficients, the
high-amplitude QDC cuts provide only a slight improvement of 2-5 ps.
- Table 4 contains a summary of the time resolution results for each run. The σ from the fit
represents the convolution of all six PMTs in the system. The reference counter (scintillator bar)
resolution is obtained by σm = sqrt(σT2
- 0.5*σr2), which is the value shown in
Table 4 with σr = 58 ps.
Table 4: LASPD bar timing resolution results using three bar method with Hamamatsu R9779
PMTs
run#
|
Raw Resolution [ps]
|
Time Walk Corrected Resolution [ps]
|
Time Walk Corrected with QDC cuts Resolution [ps]
|
Time Walk Corrected with High-amp QDC cuts Resolution [ps]
|
1496
|
T Spectrum: 314
|
149
| 130
|
128
|
1498
|
T Spectrum: 284
|
110
|
88
|
83
|
January 6th - 9th, 2015
- Time resolution measurements using 3-bar tests with R9779 PMTs Part 2:
Using the EJ-200 plastic scintillators (5 cm x 5 cm x 30 cm) with each end coupled with
optical grease to a R9779 PMT for readout.
The HVs for the six PMTs were approximately gain matched.
- For the set of measurements, the bars were stacked vertically with respect to each other.
The three trigger bars were oriented parallel to each other. The vertical spacing is about
24.2 cm between the top of the top bar and the top of the bottom bar. The trigger was generated
by forming an "AND" of the left and right PMT signals of the top and bottom scintillator bars.
The top right bar's signal was delayed by 16 ns with respect to the other signals so that this PMT
would always carry the trigger time. The middle bar was not part of the trigger.
- For all measurements, the PMT signals were split from the dark box with one signal passed to a
discriminator channel set at -50 mV. The other signal was sent through 80 feet (123 ns)
of delay cable before being sent to the QDC module.
- For run 1480, the PMT HVs were set modestly low. For run 1484 the HVs were increased by about 100 volts, and
for run 1487, the HVs were increased by approximately another 100 volts. However, when the HVs were adjusted to
better match the the MIP response, the HV channels were incorrectly adjusted, which caused large differences
in most of the MIP values (from 525 up to 900 QDC channels) for all six PMTs.
- Conclusions and Observations:
- The timing resolution of the three bar measurments was determined using the procedure discussed in the
CLAS12 FTOF report.
- For each run, the QDC pedestals were deterimined in order to subtract the offset from the QDC response. The
pedestal was isolated by placing an anti-cut on the TDC for the corresponding QDC channel. Table 1 summarizes
the pedestal and MIP response for each PMT.
Table 1: Timing resolution data using three bar method with Hamamatsu R9779
PMTs
run#
|
Discriminator threshold [mV]
|
Pedestals
|
measured MIP positions
|
comments
|
1480
|
-50
|
file1
|
file2
|
Trigger: "AND" of four pmts
top and bottom bars |
1484
|
-50
|
file3
|
file4
|
Trigger: "AND" of four pmts
top and bottom bars HV increased to move MIP to ~ 450 channels |
1487
|
-50
|
file5
|
file6
|
Trigger: "AND" of four pmts
top and bottom bars HV increased to move MIP to ~ 600-700 channels |
- Then the time from the TDC was plotted versus the pedestal subtracted QDC response (1/sqrt(QDC - pedestal)): runs
1480, 1484
and 1487. The top right PMT is self-timed, and hence there
is no time dependence wrt the QDC signal.
- A profile histogram was created so that the distribution could be fit: runs
1480, 1484
and 1487. The slope from the fit was used to correct
for the time-walk effect.
- Interestingly, the data from run 1487, do not show any significant timewalk effect. Updated on January 13:
It was found that the TDC and QDC information from the two modules were not correlated with each other. This
can be easily seen by cutting on the TDC. The pedestal should be completely removed, but it is not for at least
runs 1487 and 1494. More than likely this is a DAQ problem. One or both modules during the run have extra words
for one or more events. Since this data is not cleared before the next event, the TDC and QDC information fall
out of synchronization with each other. The CODA crl file was updated to clear any extra words that are found
before moving on to process the next event.
- A cut was also placed on the pedestal subtracted QDC distributions to remove the low energy
events. For run 1480, the cut was placed at channel 153 for each PMT signal.
For run 1484, the following QDC cuts were chosen:
- Top right cut = 300 channels
- Top left cut = 350 channels
- Middle right cut = 350 channels
- Middle left cut = 350 channels
- Bottom right cut = 360 channels
- Bottom left cut = 360 channels
For run 1487, the following QDC cuts were chosen:
- Top right cut = 430 channels
- Top left cut = 750 channels
- Middle right cut = 500 channels
- Middle left cut = 440 channels
- Bottom right cut = 670 channels
- Bottom left cut = 560 channels
- Using the expression T = 0.5*(tt + tb) - tm, where tt, tm, and
tb represent the time the particle passes through the top, middle, and bottom bars, respectively, the time
resolution of the bars was determined.
- The time can be expressed in terms of the times of PMT signals as T = 0.25*(ttr + ttl
+ tbr + tbl) - 0.5*(tmr + tml), where the subscript 'r' and 'l' denote the
right-side and left-side pmts, respectively, for that particular bar.
- The time resolution is then expressed as σref = sqrt(2/3)* σT, where σT
is determined from a fit to the timing distribution.
- Added on January 27: The initial time-walk slopes were then optimized by varying one slope and holding
the other 5 fixed until the timing resolution was minimized. This procedure was repeated for each PMT.
- The timing resolution for the two runs (1480 and 1484) are quite different from each other, which indicates there is a
dependence on PMT gain. Without applying QDC cuts to
remove the low energy plateau, a tail is still seen even after the time-walk corrections are applied:
1480 and
1484, though the timing resolution does improve by about
50% and 100%, respectively, with the time-walk correction applied.
- By applying QDC cuts, the long tail is greatly reduced in both the distribution
after time-wallk corrections: 1480 and
1484. The QDC cut was not applied to the left-hand side
plots, in which the time walk correction is also not applied.
- For run 1487, there is no difference when the timewalk correction
is applied. Finally there is only a slight improvement when the
QDC cuts are applied. Clearly the timing distributions are not gaussian,
and the fit quality is poor.
- Table 2 contains a summary of the time resolution results for each run. The σ from the fit represents the
convolution of all six PMTs in the system. The reference counter (scintillator bar) resolution is obtained by multiplying σ
by sqrt(2/3), which is the value shown in Table 2.
Table 2: Timing resolution results using three bar method with Hamamatsu R9779
PMTs
run#
|
Raw Resolution [ps]
|
Time Walk Corrected Resolution [ps]
|
Time Walk Corrected with QDC cuts Resolution [ps]
|
1480
|
287
|
T Spectrum: 184
|
T Spectrum: 169
|
1484
|
T Spectrum: 202
|
66
|
54-58
|
1487
|
T Spectrum: 178
|
N/A
|
N/A
|