Preshower Tile Test at UVa, 2014
Summary and Next step:
- Test thicker fibers or fiber with higher dye concentration.
- Will setup COMPASS module, vertically but will be tilted (dark box not tall enough), test MIP response;
- Setup TDC and test timing response;
- Test R11102 and R9800 PMTs;
- Current best estimate for Preshower: 50*1.05(glue or grease)*0.8(connector)*(0.63-0.83)=26 for LA, 35 for FA
December 8th - 18th, 2014
- Time resolution measurements using 3-bar tests:
Using the EJ-200 plastic scintillators (5 cm x 5 cm x 30 cm) with each end coupled with
optical grease to a XP2262 PMT for readout. The HVs for the six PMTs were approximately
gained match so the MIP would be near channel 1100.
- For the first set of measurements, all three bars were resting on the bottom of the dark box, and
the trigger was generated by forming an "AND" of the left and right PMT signals. Finally these
"AND" signals were "ORed" together, so that the trigger was the "OR" of the "ANDs". For each bar,
the right PMT's signal was delayed by 16 ns using a ORTEC delay box (model DB463).
The three trigger bars were oriented parallel to each other.
- For the second set of measurements, a new
stand was built so that the
bars could be stacked vertically with respect to each other. The three trigger bars were oriented
parallel to each other. The vertical spacing is about 24.2 cm between the top of the top bar and
the top of the bottom bar. The trigger was generated by forming an "AND" of the left and right PMT
signals of the top and bottom scintillator bars. The top right bar's signal was delayed by 16 ns
with respect to the other signals so that this PMT would always carry the trigger time. The middle
bar was not part of the trigger.
- For all measurements, the PMT signals were split from the dark box with one signal passed to a
discriminator channel set at either -40 mV or -80 mV. The other signal was sent through 80 feet (123 ns)
of delay cable before being sent to the QDC module.
- Conclusions and Observations:
- With only one bar in the trigger, the MIP peak sits on top of a large backward, which is suppressed in
the vertical bar test. After run 1462, the threshold was doubled to -80 mV to suppress the low energy "peak"
near caused by the discriminator threshold. In the three bar vertical measurements, the low energy events
show up as a plateau below the MIP location.
- With the different trigger configurations, the pedestal moved around by about 10-15 channels. Runs 1465
and 1466 were taken to find the pedestals for the PMTs of the top and bottom bars, which cannot be seen in
the normal running configuration due to these bars being in the trigger.
- The energy deposited by the cosmics is attenuated by the shelves of wood that the bars rested and the other bars
by about 11% on average from the top to the bottom bars. The thickness of the plywood is 0.75 inches (1.9 cm).
- The timing resolution of the three bar measurments was determined using the procedure discussed in the
CLAS12 FTOF report.
- For each run, the QDC pedestals were deterimined in order to subtract the offset from the QDC response. The
pedestal was isolated by placing an anti-cut on the TDC for the corresponding QDC channel.
- Then the time from the TDC was plotted versus the pedestal subtracted QDC response (1/sqrt(QDC - pedestal)): run
1464 and
1467. The top right PMT is self-timed, and hence there
is no time dependence wrt the QDC signal.
- A profile histogram was created so that the distribution could be fit: run
1464 and
1467. The slope from the fit was used to correct
for the time-walk effect.
- In a second iteration, a cut was also placed on the pedestal subtracted QDC distributions to remove the low energy
events:
- Top right cut = 800 channels
- Top left cut = 1050 channels
- Middle right cut = 750 channels
- Middle left cut = 750 channels
- Bottom right cut = 780 channels
- Bottom left cut = 700 channels
Applying these cuts was seen to greatly improve the timing resolution.
- Using the expression T = 0.5*(tt + tb) - tm, where tt, tm, and
tb represent the time the particle passes through the top, middle, and bottom bars, respectively, the time
resolution of the bars was determined.
- The time can be expressed in terms of the times of PMT signals as T = 0.25*(ttr + ttl
+ tbr + tbl) - 0.5*(tmr + tml), where the subscript 'r' and 'l' denote the
right-side and left-side pmts, respectively, for that particular bar.
- The time resolution is then expressed as σref = sqrt(2/3)* σT, where σT
is determined from a fit to the timing distribution.
- The timing resolution for the two runs (1464 and 1467) are consistent with each other. Without applying QDC cuts to
remove the low energy plateau, a long tail is still seen even after the time-walk corrections are applied:
1467, though the timing resolution does improve by about
50% with the time-walk correction.
- By applying QDC cuts to the tdc signals in the expression for "T", the long tail is greatly reduced in both the distribution
before and after time-wallk corrections: 1464 and
1467.
- As a last iteration, the QDC cuts were also applied when the time-walk correction was determined. With better time-walk
coefficients, the timing resolution improved an additional 17%:
1467. The limitation can be traced to two factors: a non-gaussian tail around 2.4 ns and above and substructure that is
seen in the time versus QDC plots near the MIP location. This substructure is not removed by the time-walk corrections and
seems to be isolated to only the left-hand side pmts for each bar.
- This text file contains a summary of the time resolution
results.
Timing resolution using three bar method with XP2262
PMTs
run#
|
Discriminator threshold [mV]
|
Pedestals
|
measured MIP positions
|
comments
|
1462
|
-40
|
file1
|
file2
|
Trigger: "OR" of three bars
|
1463
|
-80
|
file3
|
file4
|
Trigger: "OR" of three bars
|
1464
|
-80
|
file5
|
file6
|
Trigger: "AND" of four pmts
top and bottom bars |
1465
|
-80
|
file7
|
----
|
Trigger: "AND" of two pmts
only top bar |
1466
|
-80
|
file8
|
----
|
Trigger: "AND" of two pmts
only bottom bar |
1467
|
-80
|
file9
|
file10
|
Trigger: "AND" of four pmts
top and bottom bars |
October 16th - November 13th, 2014
- Test PMT-fiber-tile time resolution:
Using the hexagon IHEP tile with 2x 1-mm diameter fibers coupled to a XP2262 PMT
for readout. The HV for the PMT viewing the WLS fibers was kept at -2100 V (HV3) to
compare with earlier tests using the same tile and PMT.
- For these measurements, the trigger was generated by two scintillator bars from
A. Camsonne. Three of the PMTs are similar to XP2262, and the other is a XP2972.
The top trigger bar consists of one XP2262 and the other is the XP2972. The bottom
bar is readout using two XP2262's. The two trigger bars were oriented parallel to each other.
- The HV settings were adjusted to ensure each PMT had average pulse heights that would
pass a -30 mV threshold and were approximately the same amplitude to each other. The
HV settings are HV1 = -1425 V, HV2 = -925 V, HV4 = -1525 V, and HV5 = -1325 V.
- The discriminator thresholds were changed from -100 mV to -30.5 mV.
- The delay for the PMT pulse to QDC was changed to 20 ns to avoid
having the pulse too close to the closing edge of the gate. This change was
made using the PS 792 delay module.
- All four trigger PMTs signals are sent to the TDC after discrimination for timing
studies.
- The optical grease on the PMT surface was smoothed out before placing the WLS fiber
ends against the PMT surface.
- The fibers were also repositioned into the tile groove, since the outer turns had come out
of the groove. No additional optical grease was applied.
- The first test is a reproducibility study to ensure that the measurements are comparable
to the MIP response achieved in the earlier study.
- Conclusions and Observations:
- The series of tests described below indicates some reproducibility issues in the system, which
probably indicates a potential instability somewhere.
- The MIP response compared to run 1384, which should be a similar configuration, was found have
a 25% increase (2598/2074). At this point, it is unclear what has caused the increase in amplitude,
though perhaps the repositioning of the fibers and smoothing of the grease increased the light
collection efficiency.
- The SPE was never clearly identified, and the number of p.e.is unknown, since the MIP is quite
different than expected.
- To better identify the SPE, the trigger was changed from the "AND" of all four PMTs to be the "AND" of the
top-right PMT and the bottom-left PMT (one set of diagonal PMTs were chosen), i.e., a looser trigger was
formed, which allows lower energy events to form a trigger. However, the SPE was not seen in this configuration
either.
- Run 1444 was taken with the splitter to send the tile timing information to the TDC. Interestingly, the MIP
is comparable to runs 1384 and 1385, which should be similar testing conditions.
- With succesive runs, the MIP value continued to increase. Between runs 1445 and 1446, it was realized that
tile PMT might be viewing light escaping from the tile through holes in the tyvek wrapping. Before run 1448,
the tile was removed and the tyvek paper was resecured. Also the tile PMT was covered with black material to
avoid any unwanted light from entering the PMT.
- For run 1448, the MIP was still about 15% higher than before, but the SPE was identified and the number of p.e. is
close the the maximum achieved for run 1384.
Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
with optical grease in grooves
run#
|
# fiber turns
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using single p.e. position
|
comments
|
1442
|
2.5 each fiber
|
SPE?
|
2598
|
----
|
Trigger: four pmts
|
1443
|
2.5 each fiber
|
----
|
2468
|
----
|
Trigger: two pmts (diagonal)
|
1444
|
2.5 each fiber
|
----
|
1042
|
----
|
Trigger: four pmts
Using Splitter to TDC |
1445
|
2.5 each fiber
|
----
|
2894
|
----
|
Trigger: four pmts
Without Splitter |
1446
|
2.5 each fiber
|
----
|
3093
|
----
|
Trigger: four pmts
Without Splitter |
1447
|
2.5 each fiber
|
----
|
937
|
----
|
Trigger: four pmts
Using Splitter to TDC |
1448
|
2.5 each fiber
|
35.2 ± 2.3
|
2384
|
67.8
|
Trigger: four pmts
Without Splitter |
October 3rd - 7th, 2014
- Test SDU Square SPD tiles:
Using a 0.5 cm thick, SDU square tile with a XP2262 PMT.
The HV for the PMT viewing the tile was kept at -1850 V to compare with
the tests on the IHEP hexagon 2 cm thick tile.
- The delay for the PMT pulse to QDC was changed from 44 ns to 32 ns to avoid
having the pulse too close to the closing edge of the gate. This change was
made using the PS 792 delay module.
- All runs for this set of tests were taken without the 50-50 splitter.
- The tile was wrapped in tyvek paper and then wrapped with black tape, except for
a 5 cm opening on one end. Optical grease was placed on the tile surface and then
was placed against the PMT cathode.
- Conclusion:
- The SPE for run 1430 is right next to the pedestal and sitting on the pedestal's tail.
So even though the fit error is small, the systematic is larger than half a QDC channel.
- The SPE is easily identified in run 1432 and approximately where expected from earlier data
with WLS fibers embedded.
- The NPE value of 48 is considerably more accurate than the value of 65.
- The gain factor for HV = -1850 V and HV = -2100 is determined to be (1579/333) = 4.7, which is
significantly larger than the estimated factor of 3.4 from the XP2262B specficiations. This changes
the NPE estimate for the IHEP 2 cm hexagon tile direct coupling measurement from earlier.
- Run 1433 with the 3 mm tile had the tyvek paper printed side facing the tile surface. Though
only half of the paper had print, this caused a substantial drop in signal amplitude seen in
the QDC (~265) versus the expected 945.
- Run 1438 has the 3 mm tile rewrapped with tyvek paper with the printed side facing away
from the tile surface. At worse, I would expect the same result as run 1433, but instead found
an additional drop of about 30% in amplitude, while the SPE remained in the same position.
This indicates an additional inefficiency in light collection compared to the earlier measurement.
MIP Response with SDU Square Tile, No Fibers
run#
|
measured single p.e. position
|
measured MIP
position
|
measured # p.e. using single p.e. position
|
comments
|
1430
|
5.1 ± 0.5
|
332.8
|
65.2
|
XP2262, HV = -1850 V 5 mm tile
|
1432
|
32.8 ± 1.4
|
1579
|
48.1
|
XP2262, HV = -2100 V 5 mm tile
|
1433
|
32.4 ± 0.5
|
264.8
|
8.2
|
XP2262, HV = -2100 V
3 mm tile, printed side of tyvek is facing the tile
|
1436
|
----
|
2361
|
----
|
XP2262, HV = -1800 V
2 cm IHEP hexagon tile
|
1438
|
32.0 ± 0.6
|
203.1
|
6.3
|
XP2262, HV = -2100 V
3 mm tile, rewrapped with printed side facing out
|
September 26th - October 2nd, 2014
- Test PMT-tile time response:
Using IHEP hexagon tile with wide grooves and either a XP2262 PMT or R9800 PMT.
The HV for the PMT viewing the tile was reduced to -1850 V to avoid
saturating the QDC. The estimated reduction in gain is a factor of
3.37 as determined from the gain vs. HV curve.
- The delay for the PMT pulse to QDC was changed from 12 ns to 32 ns to avoid
having the pulse too close to the opening edge of the gate. This change was
made using the PS 792 delay module.
- The tile was wrapped in tyvek paper and then wrapped with black tape, except for
a 5 cm opening on one end. Optical grease was placed on the tile surface and then
the PMT was placed against the open end. It should be noted that there is a very small
region where the tile and PMT do not overlap, since the tile is rectangular
and the PMT circular.
- Conclusion:
- The rise time of the PMT pulse was determined using an oscillscope and taking the time
difference between the time at 10% and 90% of the signal pulse height.
- The rise time from the XP2262 PMT specifications is 2-2.3 ns,
depending on the voltage divider, and 1.0 ns for the R9800.
- From the measured rise times, it appears the rise time is dominated by the scintillator material
rise time of ?? ns.
- The trigger formed from the "AND" of the top and bottom CSI crystals has a
time resolution of 19 ps based on the fit. The trigger going into the TDC is self-timed, since the
stop for the TDC is also generated by the trigger. This explains the great timing resolution for the
trigger. The smallness of the number (19-20 ps) compared to the TDC resolution is an artifact of
histogram binning.
- The time resolution of the tile signal is about 630 ps for both types of PMT, though the
peak has long tails. For the first part of these tests, the tile was viewed by only
one PMT, and the position dependence of the hit on the tile cannot be removed.
- During run 1413, it became clear that the pedestal was still wide compared to few channels
previously seen without the splitter. In fact, the pedestal is about 22 channels wide now, which
makes it difficult to find the spe peak. After some lengthy debugging, it was found that the wide
pedestals were caused by a series of connectors leading to the QDC input. I replaced these
BNC-type connectors with a single lemo barrel connector. The pedestal width is now a few
channels without going through the splitter.
- Runs 1413 and 1425 are attempts to find the spe to determine the number of photoelectrons, though the spe
was not identified.
- Run 1427 used the R9800 to readout the tile with two 1-mm diameter fibers and grease in the grooves. This
run can have an approximate direct comparison with run 1411, which involves the direct light collection from
the tile without fibers. There is a factor of four difference (951.5/237.6). However, the light collection
for run 1413 might be reduced by having the empty circular groove in the tile.
- Due to the low gain for the R9800, the SPE is
not visible. With the same tile and a XP2262 PMT, the SPE is at
channel 29 with about 70 photoelectrons.
The R9800 and XP2262 have about the same QE (25-26%) at 400 nm, taking the ratio of amplitudes and assuming the same npe,
then the SPE should be around channel 3.4, which cannot be isolated from the pedestal.
-
Timing Resolution and MIP Response with IHEP Tile, No Fibers
run#
|
Rise Time [ns]
|
Measured MIP Position
|
Estimated #pe
|
Trigger Time Resolution [ps]
|
Tile Time Resolution [ps]
|
Time Walk corrected* [ps]
|
comments
|
1407
|
2.76
zoomed in image
|
1353
|
387-436a
|
19
|
631
|
570
|
HV = -1850 V; SPE not located;
XP2262
|
1410
|
N/A
|
159
|
N/A
|
N/A
|
N/A
|
N/A
|
HV = -1100 V; Gain too low;
R9800
|
1411
|
2.48
zoomed OUT image
|
472.3
|
---- |
19
|
625
|
522
|
HV = -1350 V; SPE not located
R9800
|
1412
|
----
|
589.6
|
---- |
19
|
616
|
----
|
HV = -1420 V; SPE not located;
R9800
|
1413
|
----
|
951.5
|
---- |
19
|
N/A
|
N/A
|
HV = -1350 V; Splitter Bypassed;
R9800
|
1425
|
----
|
936.2
|
---- |
20
|
N/A
|
N/A
|
HV = -1350 V; Splitter Bypassed;
R9800
|
1427
|
----
|
237.6
|
---- |
---
|
N/A
|
N/A
|
HV = -1350 V; Splitter Bypassed;
R9800 readout of tile with two fibers.
|
1428
|
----
|
342.1
|
---- |
---
|
N/A
|
N/A
|
HV = -1450 V; Splitter Bypassed;
R9800 readout of tile with two fibers.
|
aThe number of photoelectrons was estimated by taking the reduction
in gain factors: 2 for the splitter, and
4.7 as determined on
October 3 for the change in HV.
The total gain reduction for the combined changes is
9.4. The spe was located around channel 29.2 from
the tests with
two fibers or 32.6 from the 5 mm tile measurements. Dividing the spe channel
by the reduction due to the changes gives 3.1-3.5 channels. npe = MIP/SPE = (1353/3.1) = 436 photoelectrons
or (1353/3.5) = 387 photoelectrons.
*The time walk correction was applied by plotting the TDC time
in nanoseconds versus 1/sqrt(ADC - pedestal) and
fitting
the dependency to a line. This correction was then applied to the raw time using the expression:
timecorr = timeraw - timetw,
where
timetw = a/sqrt(QDC - pedestal)
with a being the slope of the line, and QDC is the QDC value for each event.
September 25th, 2014
- TDC time resolution calibration:
Using the only the top CSI crystal as the trigger. HV stayed the same
- The trigger pulse was delayed going to the TDC by selecting different
delays from the Phillips Scientific Dual Delay module. Only the bottom
channel was used. Unfortunately, I could not easily measure the other
half of the TDC range, since I was unable to easily remove delay between
the pulse and the TDC stop using the same test setup.
- The time delay was measured on the lab oscilloscope with a estimated
uncertainty of 0.2 ns. Then the TDC channel was measured with runs taken
for the various delays listed in the table below.
- Conclusion:
- The splitter was verified to divide the signal 50-50. Hence, any data taken
going through the splitter has its amplitude reduced by a factor of 2.
- For run 1401, using the maximum possible delay per channel of the PS 792, no signal
was seen in the TDC. No attempt was made to understand this lack of signal.
- The correlation between the added delay and the TDC channel shows a reasonable linear
relationship.
More than
likely the error estimate of 0.2 ns is over estimated.
- The TDC resolution was determined by taking TDC channel without delay (run 1391) as the
reference channel and subtracting the results from this reference.
- The measured delay time from the oscilloscope in picoseconds was then divided by the
channel difference:
resolutioni = (Osc. delayi)/(1962 - TDCi),.
where i represents the run number.
- The mean of the 9 measurements was determined to provided a
TDC resolution = 36.2 ± 0.6 ps.
-
-
Calibration of CAEN V775 TDC Resolution
run#
|
TDC Channel
|
Peak Width
|
Delay [ns]
|
oscilloscope measured delay [ns]
|
comments
|
1391
|
1962
|
0.56
|
0
|
N/A
|
Baseline Delay to TDC
|
1394
|
1887
|
0.6
|
0
|
2.8
|
Delay through PS 792
|
1395
|
1444
|
0.6
|
16
|
18.6
|
Delay through PS 792 + 16 ns
|
1396
|
1223
|
0.59
|
24
|
26.6
|
Delay through PS 792 +24 ns
|
1397
|
999
|
0.57
|
32
|
34.8
|
Delay through PS 792 + 32 ns
|
1398
|
778
|
0.57
|
40
|
42.8
|
Delay through PS 792 + 40 ns
|
1399
|
555
|
0.57
|
48
|
50.8
|
Delay through PS 792 + 48 ns
|
1400
|
335
|
0.59
|
56
|
58.8
|
Delay through PS 792 + 56 ns
|
1401
|
N/A
|
N/A
|
63.5
|
66.0
|
Delay through PS 792 + 63.5 ns
|
1402
|
1778
|
0.68
|
4
|
6.7
|
Delay through PS 792 + 4 ns
|
1403
|
1558
|
0.79
|
12
|
14.6
|
Delay through PS 792 + 12 ns
|
September 12th-September 24th, 2014
- Multi-fiber test: Using the
IHEP Preshower hexagon tile. Now use two 1.0-mm diameter, 1.5-m long Y11
fiber. The fiber ends were cut off so that only 20 cm of fiber is not
embedded in the tile. The fiber ends were then polished with polishing
paper from JLab. HV stayed the same
- Optical grease was coated onto the pmt cathode surface, and the four fiber
ends were placed up against the coated cathode. The fibers were held rigid using
a plastic block from the machine shop.
- Conclusion:
- The first test produced a lower than expected number of photoelectrons,
especially considering the shorter fiber length. One issue is that the
quality of the fiber ends was not assessed after placing the fibers
through the narrow holes in the plastic block.
- The fiber ends were repolished, and the test was conducted again with the
newly repolished ends. However, only a slight improvement is seen, which is
not statisically significant.
- After the first run, the position of the s.p.e. appears to have shifted lower,
though it is within 2 sigma from the average.
- After the aluminized mylar test, it was noted that the inside of the top panel
for the mylar is discolored, and the mylar was bulging out, which might have caused
light loss reducing the yield for this test.
- The table below includes the data with different turns for each of the two fibers.
-
Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using average single p.e. position:
28.9 |
comments
|
1371
|
2.5 each fiber
|
155.4
|
30.2 ± 1.0
|
1418 |
49.1 |
Optical grease on PMT Tyvek wrapping
|
1372
|
2.5 each fiber
|
155.4
|
27.5 ± 1.3
|
1429
|
49.4
|
Fiber ends repolished Tyvek wrapping
|
1373
|
2.5 each fiber
|
155.4
|
28.3 ± 1.4
|
1345
|
46.5
|
Aluminized mylar wrapping
|
1374
|
2.5 each fiber
|
155.4
|
26.8 ± 1.9
|
1479
|
51.2
|
Reproducibility of Run 1372
|
1375
|
1.5 + 2.5 fibers
|
120.1
|
27.8 ± 1.0
|
1324
|
45.8
|
Bottom fiber: 2.5 turns;
Top fiber: 1.5 turns
|
1376
|
1.5 fibers each
|
91.8
|
27.4 ± 1.3
|
1048
|
36.3
|
Bottom fiber: 1.5 turns;
Top fiber: 1.5 turns
|
1378
|
0.5 + 1.5 fibers
|
63.5
|
29.5 ± 0.5
|
970.5
|
33.6
|
Bottom fiber: 1.5 turns;
Top fiber: 0.5 turns
|
1379
|
0.5 fibers each
|
35.3
|
29.4 ± 1.2
|
630.2
|
21.8
|
Bottom fiber: 0.5 turns;
Top fiber: 0.5 turns
|
- Now repeat the steps above but with optical grease in the tile grooves. The grease is
Saint-Gobain BC-630, and was applied with a thin plastic cable tie.
- Conclusion:
- The addition of optical grease improves the yield on average by a factor of about 1.4.
- The variation in improvement is likely due to how well the grease is applied. For example, the
lower improvement for less fiber turns is probably due to difficulty in placing the grease deep into
the narrow grooves. For the 2.5 turns per fiber, the last turn of the top fiber is difficult to
keep inside the groove, but this is true with or without grease.
- For run 1385, the PMT pulse was sent through the splitter, where half was taken to
QDC and the other half to a discriminator (-40 mV threshold) and to the TDC. For run 1386, the splitter
was bypassed and the pmt pulse passed directly to a discriminator and then to the TDC.
- The full scale range (FSR) was set to 850 ns (212.5 ps/ch resolution). However, the timing
difference between the trigger and stop and the tile signal and stop were measured on the oscilloscope
to be 182 ns and 218 ns, respectively. If a TDC resolution of 212.5 ps/ch is used, then the time
measured by the TDC is a factor of 2 too large. The oscilloscope measured values indicated that the
resolution is really about 106.25 ps/ch and the FSR = 425 ns. It's not clear why this is. When
CODA is downloaded, it claims that the FSR is set to 850 ns. The calculation in the JLab V775 library
appears to be correct. An independent TDC resolution measurement needs to be performed to clarify the
actual resolution, though no further measurements are planned at FSR = 850 ns.
- The tile timing resolution does appear to be affected by the splitter. Without passing through
the splitter, the timing resolution is about 100 ps better.
- The difference in trigger time resolution is probably due to the quality of the fit and is probably closer
to 78 ps.
Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
with optical grease in grooves
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using average single p.e. position:
29.2 |
Ratio of
Grease to no Grease Result
|
comments
|
1380
|
0.5 fibers each
|
35.3
|
29.5 ± 0.6
|
843.4
|
28.9
|
1.33
|
Bottom fiber: 0.5 turns;
Top fiber: 0.5 turns
|
1381
|
0.5 + 1.5 fibers
|
63.5
|
29.4 ± 2.4
|
1337
|
45.8
|
1.36
|
Bottom fiber: 1.5 turns;
Top fiber: 0.5 turns
|
1382
|
1.5 fibers each
|
91.8
|
29.0 ± 0.6
|
1673
|
57.3
|
1.58
|
Bottom fiber: 1.5 turns;
Top fiber: 1.5 turns
|
1383
|
1.5 + 2.5 fibers
|
120.1
|
27.0 ± 2.2
|
1951
|
66.8
|
1.46
|
Bottom fiber: 2.5 turns;
Top fiber: 1.5 turns
|
1384
|
2.5 each fiber
|
155.4
|
28.5 ± 2.2
|
2074
|
71.0
|
1.39
|
Bottom fiber: 2.5 turns;
Top fiber: 2.5 turns
|
1385
|
2.5 each fiber
|
155.4
|
?
|
1019
|
69.8a
|
N/A
|
Using Splitter to TDC
FSR = 850 ns
|
aThe average single p.e. position divided by 2 was used.
Timing Resolution with Multi-fiber test results (two Y11, 1mm dia, 1.177m length)
with optical grease in grooves
run#
|
# fiber turns
|
Measured MIP Position
|
Trigger Time Resolution [ps]
|
Tile Time Resolution [ns]
|
comments
|
1385
|
2.5 each fiber
|
1019
|
52
|
1.117
|
Using Splitter to TDC
FSR = 850 ns
|
1386
|
2.5 each fiber
|
N/A
|
78
|
1.013
|
No Splitter to TDC
FSR = 850 ns
|
September 5th-September 9th, 2014
- 2-mm diameter fiber test: Using a
IHEP Preshower hexagon tile with wider grooves wrapped in Tyvek. Now using one
2.0-mm diameter, 3-m long Y11 (200 ppm) fiber. HV stayed the same. The gooves
are 7.5mm deep, with diameter 9cm and each side "leg" approx. 3.5cm long towards
the end of the tile.
- The yields are much lower than we expected-- if we assume the same bending loss
and attenuation length as the 1mm fiber. (For example, the test gave 25 p.e. for
5.5 turns). However, a bending loss in the 25%-per-turn range can explain the yield
well. And this is consistent with the 2mm fiber being non-S type, and our grooves
have 9cm diameter.
- Additionally, as the fiber was unwound to reduce the number of embedded turns,
it was noticed that there are several cracks along the fiber where it was wound into
the tile. This probably explains the cause for the lower yield.
-
2-mm diameter fiber test results (one Y11 (200), 2 mm dia, 2.0 m length), no grease in grooves
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using average single p.e. position |
#p.e. per fiber length (cm-1)
|
# p.e. expected if using minimum lengthb |
1352
|
5.5
|
162.5
|
34.9
|
862 |
25.0 |
0.154
|
~
|
1353
|
4.5
|
134.2
|
33.5
|
814.6
|
23.7
|
0.176
|
~
|
1354
|
3.5
|
106.0
|
34.8
|
781.4
|
22.7
|
0.213
|
~
|
1355
|
2.5
|
77.7
|
35.0
|
787.1
|
22.9
|
0.295
|
~
|
1356
|
1.5
|
49.4
|
34.4
|
722.5
|
21.0
|
0.425
|
~
|
1357
|
0.5
|
21.1
|
33.6
|
451.8
|
13.1
|
0.621
|
~
|
aThe average single p.e. position is 34.4
bminimum length is embeded length + 20cm on each side (40cm total), using 3.5m as decay length
September 3rd-September 5th, 2014
- 1-mm diameter fiber test: Back to IHEP Preshower hexagon
tile with 1.0-mm diameter, 3-m long Y11 fiber. The gooves are 7.5mm deep, with diameter 9cm and
each side "leg" approx. 3.5cm long towards the end of the tile.
- The results should be compared against those from run 1300, which found 37.7 photoelectrons;
this represents a 17% reduction in the number of photoelectrons from earlier studies. The fiber
ends had dark spots, and it was decided to polish the ends to check for improvement.
- The fiber ends were polished with 3 micron polishing paper from JLab. After polishing, 35.2
photolelectrons were observed, indicating that the low number of photoelectrons were due to poor
light collection from the darkened fiber ends.
Tyvek reproducibility test results (Y11, 1mm dia, 3m length), no grease in grooves
(HV = -2100 V versus -2105 V from before)
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using average single p.e. position a |
#p.e. per
fiber length (cm-1)
|
1350
|
6.5
|
190.8
|
32.7
|
1018
|
31.1
|
0.163
|
1351
|
6.5
|
190.8
|
32.7*
|
1150
|
35.2
|
0.184
|
*The single p.e. position was not identified for run 1351,
so the value from run 1350 was used instead.
August 4th-September 2nd, 2014
- Shashlyk COMPASS-II Module test: Using the same CSI
blocks and pmts for triggers as in the preshower tile tests. HV stayed the same for all runs,
except the last, where the gain was decreased by a factor of four on the shashlyk PMT.
- Conditions:
- For most of the test runs, the shashlyk module was oriented
vertical,
except where noted
as horizontal.
- The top CsI crystal was located on the top shelf of the dark
box, and the bottom CsI crystal was located on the lower shelf
of the box. The vertical separation between the two crystals
was about 30 inches (~76 cm). Most of the time, the crystals
were oriented perpendicular to each other to narrow their area
to better match the area of the shashlyk module.
- The overlap area of the crystals was 5 x 5 cm2,
and the lateral area of the shashlyk is 3.8 x 3.8 cm2.
- The timing of the shashlyk signal wrt the gate to the QDC was
adjusted by adding 20 ns of delay using a 4-ns lemo cable and 16 ns
from a PS 792 delay module.
- Run 1331 and runs from 1345 have the splitter removed from the
shashlyk cabling. For other runs, the shashlyk signal is split
and sent to the QDC after delay and also to a discriminator and
channel 0 of the v775 TDC.
- From run 1348, the CsI trigger thresholds were set to -100 mV,
previously set to -40 mV to remove low energy triggers.
- Conclusions:
- With the signal passing through the splitter, it was found
that the shashlyk signal did not produce saturation at high QDC
channels, run 1323.
- The splitter at UVA was found to both shift the pedestal and broaden the
pedestal by about a factor of 10, from 1-2 channels up to 10 channels wide.
- The saturation effect of the QDC was tested using the signal from top CsI
crystal into the QDC, run 1334.
The shelf above channel 3400 are events that saturated the QDC.
- Since, the MIP was hard to identify with the shashlyk vertical, some data was
taken with the shashlyk oriented horizontal and lying on top of the bottom CsI
crystal. For most of these tests, only the bottom CsI crystal formed the
trigger. From the horizontal tests, the MIP was easily identifiable,
run 1337.
- In the table below, the threshold refers to the trigger threshold either coincidence,
bottom CsI, or shashlyk. The shashlyk discriminator threshold is
typically -40 mV, unless otherwise noted. The efficiency is defined
as the number of events with a TDC hit compared to all triggered
events.
- This figure shows a comparison of a few of the different configurations:
shashlyk_horizontal_tests.jpg. The red and green curves are from run 1337 with and
without a TDC cut, respectively. The black curve curve is from run
1344, and the magenta curve is from run 1345, where the splitter has
been removed from the circuit. Removing the splitter increases the
amplitude by about a factor of two, while narrowing and shifting the
pedestal to lower channel.
- Triggering directly on the shashlyk, run 1344, indicates that the
module sees numerous events with low energy deposition. These are
eliminated by triggering on the CsI crystal below.
Horizontal Shashlyk Tests
-
run#
|
Trigger
|
Rate [Hz]
|
Trigger Threshold [mV]
|
Efficiency [%]
|
1337
|
CsI Coincidence
|
0.04 Hz
|
-40
|
N/A
|
1338
|
Bottom CsI
|
330
|
-40
|
N/A
|
1339
|
Bottom CsI |
11.5
|
-250
|
3.25
|
1341
|
Bottom CsI |
2.8
|
-500
|
16.4
|
1343
|
Bottom CsI |
2.1
|
-500 shashlyk threshold = -30 mV
|
26.6
|
1344
|
Shashlyk |
22.7
|
-40
|
N/A
|
1345
|
Bottom CsI |
2
|
-500
|
N/A
|
1346
|
CsI Coincidence |
0.01
|
-250
|
75.
|
- The vertical shashlyk tests were resumed with runs 1348 and 1349. The
CsI discriminator threshold was set at -100 mV, and the HV was decreased
from -2105 V to -1805 V between runs 1348 and 1349 for the shashlyk
PMT. For run
1348, saturation was seen above channel 3500. Reducing the gain removed the
saturated events as seen in run
1349. A comparison of the two spectra with different gains is
shown
here. Besides a cusp around channels 500-600, nothing else is
prominent.
May 22nd-May 27th, 2014
- Multi-fiber test: Using same
IHEP Preshower hexagon tile as the Tyvek test of last week. Now use two
1.0-mm diameter, 1.5-m long Y11
fiber. HV stayed the same
- Conclusion:
- Two-fiber test results are higher than expected, indicating an
attenuation length of the fiber shorter than 350cm, or a bending loss.
A rough "fit" to all tests starting 4/21 indicates that either the
attenuation length is 200cm, or the attenuation length is 350cm but
there is a 6%/turn bending loss.
- The
"# p.e. expected if using 2mm-dia fiber and mim length" (right-most
colume) of all tables starting 4/21 have been updated using 350cm
attenuation length and a 6%/turn bending loss.
- Both the expected yield using two fibers at min length and the
expected yield using 2mm fiber and min length are higher than before,
with the double-1mm-dia fiber configuration "winning" slightly. Also,
the 2mm-dia calculation assumed the same 350cm attenuation length which
is probably too optimistic. Kuraray informed me they do not make S-type
fibers (better performance in bending than non-S type) for 2mm
diameters and above. All 1mm-dia Y11 fibers we used in the test are
S-type. Which one is better awaits for further testing.
-
Multi-fiber test results (two Y11, 1mm dia, 1.5m length), no grease in grooves
run#
|
# fiber turns
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using average single p.e. position |
# p.e expected if using minimum length (two fibers)
|
# p.e expected if using 2mm dia fiber and minimum lengthc |
|
3.5 each fiber
|
N/A
|
N/A |
N/A |
~57
|
~56
|
1307
|
2.5 each fiber
|
31.1
|
1644.3
|
51.1
|
~54
|
~52
|
1308
|
1.5 each fiber
|
33.3
|
1212.5
|
37.7
|
~40
|
~37 |
May 7nd-May 21st, 2014
- Tyvek test: Back to IHEP Preshower hexagon tile with 1.0-mm diameter, 3-m long Y11
fiber. The gooves are 7.5mm deep, with diameter 9cm and each side "leg"
approx. 3.5cm long towards the end of the tile.
- Result: yield in average 10% above printer paper wrapping.
Tyvek homewrap test results (Y11, 1mm dia, 3m length), no grease in grooves
(new HV)
run#
|
# fiber turns
|
fiber length embeded (cm)
|
measured single p.e. position
|
measured MIP position
|
measured # p.e. using average single p.e. position a |
#p.e. per
fiber length (cm-1)
|
# p.e expected if using minimum lengthb |
# p.e expected if using 2mm dia fiber and minimum lengthc |
1300
|
6.5
|
190.8
|
33.8
|
1256.7
|
37.3
|
0.196
|
41.2
|
56.6
|
1299
|
5.5
|
162.5
|
33.9
|
1242.8
|
36.9
|
0.227
|
42.4
|
55.3
|
1301
|
4.5
|
134.2
|
30.9
|
1117.0
|
33.2
|
0.247
|
39.7
|
49.2
|
1305
|
3.5
|
106.0
|
34.1
|
900.1
|
26.7
|
0.252
|
33.3
|
39.2
|
1303
|
2.5
|
77.7
|
35.2
|
824.0
|
24.5
|
0.315
|
31.7
|
35.5
|
1304
|
1.5
|
49.4
|
34.2
|
579.9
|
17.2
|
0.348
|
23.3
|
24.7
|
1306
|
0.5
|
21.1
|
33.6
|
310.2
|
9.2
|
0.436
|
13.0
|
13.1
|
aThe average single p.e. position is 33.7
bminimum length is embeded length + 20cm on each side (40cm total), using 3.5m as decay length
cassuming same depth of fiber embedding (thus half embedded length) + 20cm on each side, using 3.5m as decay length
dall expected value assumed 6%/turn of light loss due to bending
May 2nd-May 6th, 2014
- Start setting up the TDC. Hardware seems to be working. Need to adjust the delay and add TDC to the analyzer.
- Aluminized mylar test: Back to IHEP Preshower hexagon tile with 1.0-mm diameter, 3-m long Y11
fiber. The gooves are 7.5mm deep, with diameter 9cm and each side "leg"
approx. 3.5cm long towards the end of the tile.
- Existing data on mylar: see (result of google search): PHENIX test on Tyvek 1055B (1997); Another BNL test (1996)-flawed in Al. Mylar;
- Summary of reflectivity info: Tyvek 1055B: 95%; Al Mylar: 90% (from more google search); copier paper: 75-80%.
- Result: yield is in average 17% above printer paper wrapping, systematic uncertainty at the 5-10% level.
aluminized mylar test results (Y11, 1mm dia, 3m length), no grease in grooves
-
run#
|
# fiber turns
|
fiber length
embeded (cm)
|
measured single p.e.
position
|
measured MIP position
|
measured # p.e. using average
single p.e. position a
|
#p.e. per
fiber length (cm-1)
|
# p.e. expected if
using minimum lengthb
|
# p.e. expected if
using 2mm dia fiber and minimum lengthc |
1280
|
6.5
|
190.8
|
25.0
|
1114
|
41.5
|
0.218
| 45.8
|
62.9
|
1281
|
5.5
|
162.5
|
26.2
|
1066
|
39.7
|
0.244
| 45.7
|
59.6
|
1282
|
4.5
|
134.2
|
27.4
|
920
|
34.3
|
0.255
| 41.0
|
50.9
|
1283
|
3.5
|
106.0
|
27.5
|
786
|
29.3
|
0.276
| 36.5
|
43.0
|
1284
|
2.5
|
77.7
|
27.2
|
668
|
24.9
|
0.320
| 32.3
|
36.1
|
1285
|
1.5
|
49.4
|
25.0
|
490
|
18.3
|
0.370
| 24.7
|
26.2
|
1286
|
0.5
|
21.14
|
28.8
|
265
|
9.9
|
0.467
|
13.9
|
14.0
|
aThe average single p.e. position is 26.81
bminimum length is embeded length + 20cm on each side (40cm total), using 3.5m as decay length
cassuming same depth of fiber embedding (thus half embedded length) + 20cm on each side, using 3.5m as decay length
dall expected value assumed 6%/turn of light loss due to bending
April 29-May 2nd, 2014
-
Test IHEP Preshower hexagon tile with 1.0-mm diameter, 3-m long BCF
fibers. The gooves are 5.5mm deep, with diameter 9cm and each side "leg"
approx. 3.5cm long towards the end of the tile.
- Embed 4.5 turns, took a movie with the scope readout to compare: BCF92 vs. Y11. Also see this directory for snapshots of MIP.
- Existing data: A quick google search found a recent super-B test showing
~60% for this comparison for straight fibers. ATLAS test (Fig.5-24 on EC TDR)
indicated that
for a bending radius of 10cm BCF91A loses >20% of light (while Y11
shows minimal loss). Our observed loss could be a combination of the
yield (absorption/emission efficiency) and the bending loss.
-
run#
|
fiber type
|
# fiber turns
|
# p.e.
|
# p.e., relative to Y11
with the same # turns
|
1267
|
BCF92
|
3.5
|
10.5
|
37% (run 1262)
|
1268
|
BCF91A
|
3.5
|
13.7
|
53% (run 1263)
|
1269
|
BCF91A |
2.5
|
10.7
|
52% (run 1264)
|
1270
|
BCF91A |
1.5
|
|
57% (run 1265)
|
April 21-28, 2014
Test IHEP Preshower hexagon tile with 1.0-mm diameter, 3-m long Y11
fiber. The gooves are 7.5mm deep, with diameter 9cm and each side "leg"
approx. 3.5cm long towards the end of the tile.
printer paper wrapping, Y11, 1mm dia, 3m length, no grease in grooves
run#
|
# fiber turns
|
fiber length
embeded (cm)
|
measured single p.e.
position
|
measured MIP position
|
measured # p.e. using average
single p.e. position a
|
#p.e. per
fiber length (cm-1)
| # p.e. expected if using minimum length b
|
# p.e. expected if
using 2mm dia fiber and minimum lengthc |
1260
|
6.5
|
190.8
|
23.4
|
997.6
|
39.6
|
0.207
| 43.7
|
60.0
|
1261
|
5.5
|
162.5
|
25.4
|
858.3
|
34.1
|
0.210
| 39.1
|
51.1
|
1262
|
4.5
|
134.2
|
25.4
|
717.7
|
28.5
|
0.212
| 34.1
|
42.3
|
1263
|
3.5
|
106.0
|
26.6
|
626.7
|
24.9
|
0.235
| 31.0
|
36.5
|
1264
|
2.5
|
77.7
|
24.7
|
510.0
|
20.2
|
0.261
| 26.3
|
29.4
|
1265
|
1.5
|
49.4
|
25.7
|
376.8
|
15.0
|
0.303
| 20.2
|
21.5
|
aThe average single p.e. position is 24.875
bminimum length is embeded length + 20cm on each side (40cm total), using 3.5m as decay length
cassuming same depth of fiber embedding (thus half embedded length) + 20cm on each side, using 3.5m as decay length
dall expected value assumed 6%/turn of light loss due to bending
March 31-Apr 9, 2014
Test Leoni fiber connector. Method: Use Chinese tile with 1.0-mm
diameter, 3-m long Y11 fiber, embed 7 turns. One end of WLS fiber is
connector to a 2-m PSM clear fiber, then clear fiber to PMT for readout.
- single WLS fiber output: nphe=469.3/24.8=20 (single p.e.; mip) <-- this is comparable to the previous result for 2 WLS fibers;
- with Leoni fiber connector, optical grease, pushed both fibers in contact, then placed gently: spectrum (no mip is observed). Played around with the connector and saw no difference;
- single
WLS -> Delrin scrap piece with a 1mm through hole -> clear fiber,
pushed both fibers in contact with a dab of optical grease, then placed
gently down. No loose movement of fiber because the hole diameter
matches fiber well: (mip). Loss of 2-m clear and Delrin connector combined: 76%, nphe ~ 14.3. Not bad! Picture of setup.
March 13-22, 2014
Test Chinese tile with 1.0-mm diameter, 3-m long Y11 fiber. The tile groove has been widened and deepened to 3x8mm.
- nearly 8 turns (run 1242) nphe=917/26.8=34.2 (single p.e.; mip), nphe per turn = 4.3;
- 7 turns (run 1244) nphe=912.1/24.9=36.6 (single p.e.; mip), nphe per turn = 5.2;
- 6 turns (run 1245,1247) nphe=788.2/24.1=32.7 (single p.e.; mip), nphe per turn = 5.5; <- note this run could have had only 5.5 turns
- 5 turns (run 1248) nphe=788.3/24.8=31.8 (single p.e.; mip), nphe per turn = 6.4;
- 4 turns (run 1249) nphe=633.7/24.4=26.0 (single p.e.; mip), nphe per turn = 6.5;
- 3 turns (run 1250) nphe=500.9/25.1=20.0 (single p.e.; mip), nphe per turn = 6.7;
- 2 turns (run 1253) nphe=378.0/23.0=16.4 (single p.e.; mip), nphe per turn = 8.2;
- 1 turn (run 1254) nphe=203.8/21.4=9.5 (single p.e.; mip), nphe per turn = 9.5;
March 5-12, 2014
Test Chinese tile with 0.5-mm diameter, 3-m long Y11 fiber:
- nearly 5 turns (run 1232) nphe=223.3/28.3=7.89 (single p.e.; mip), nphe per turn = 1.58;
- 4 turns (run 1233) nphe=191.8/26.0=7.38 (single p.e.; mip), nphe per turn = 1.85;
- 3 turns (run 1236) nphe=150.4/28.0=5.38 (single p.e.; mip), nphe per turn = 1.80;
- 2 turns (run 1238) nphe=110.8/(~29)=3.7 (single p.e.; mip), nphe per turn = 1.85;
- 1 turn (run 1239) nphe = 56.3/36.5=2.1 (single p.e.; mip), nphe per turn = 2.1;
- Data
showed that nphe yield proportional to fiber volume. With the same
volume, thicker fibers (1mm, see Feb28 run below) seems to give 15%
more light (compare to 0.5mm, 4 turns above). Based on this, using
1mm-fiber and 12 turns will give about 100 ph.e.
February 28 - March 1, 2014
Test Chinese tile:
- reduced fiber to single turns. See pictures: pic1 pic2 pic3;
- MIP response: chin_tile_20140228_singleturn.pdf or PNG; Single p.e. at about ch25. Number of p.e. about 8.5 which is what we expected if it is linear with fiber length embedded.
Summary of Chinese tile + 1.0mm fiber test:
- 2.2 turns (run 1229) nphe=409.5/26.5 = 15.5(mip), nphe per turn = 7.0;
- 1 turn (run 1230) nphe = 213/25 = 8.5(mip), nphe per turn = 8.5;
February 27, 2014
First try on the Chinese hexagon tile. For comparison:
- 1.5cm thick or 18/32", 12x12cm^2 square;
- fiber (Y11) embedded 3.5 turns (nearly 4 turns if including the straight portion);
- fiber length about 30cm on either side (see December 2013 record and picture)
- single p.e. peak around 26; MIP peak around 541; see russ_tile_20140227_3.5turns.pdf or PNG
- Chinese tile:
- 2cm thick or 24/32", 6.25-cm side hexagon;
- We
inserted a 3-m long Y11 fiber, embedded 2.3 turns (the 3rd turn keeps
popping out, only partially in the tile). The fiber sticked out pretty
long before the PMT end (at least 1m more decay length compare to LHCb
tile).
- single p.e. peak around 27;
- Based
on LHCb tile response we expected the Chinese tile MIP peak to be at:
24/18*2.3/3.5*27/26*541=492 +/- 20 given the error on determining the
s.p.e. peak.
- We measure 497 for the four hour run 1228, see Chinese-tile-20140227-4hour_2turns.pdf or PNG
- An overnight run gave slightly lower MIP response, could be because part of the fiber popped out, see pictures: pic1 pic2 pic3; MIP response: chinese-tile-20140227overnight_2turns.pdf or PNG
February 26, 2014
This is the first day we try to run the test in the new lab space (also see this pic).
Mostly tried to reproduced the earlier result on the LHCb tile. Had to
change the PMT HV (increase by 200V), replaced a PMT, etc. But at the
end got nearly 22 ph.e. for MIP.
- ;