• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name V. Sulkosky

    Log entry time 15:45:53 on December 09, 2011

    Entry number 359679

    keyword=Scaler Gates Ready for LHRS and RHRS

    Melissa, Ryan and I have setup the scaler gates for the left and right arm
    scalers.  For the most part, the cabling was in place from the spring.  We
    connected a cable from the SG_03 output from the TM module of each TS crate
    to the scaler modules (except the helicity gated modules).  
    
    The scalers on the left HRS appear to work fine.  The scalers pause and 
    clear on prestart and start counting when the DAQ is started.  At the end
    of a run, the scalers momentarily pause at the time of the last scaler 
    reading.
    
    The scalers on right HRS also appear to work fine, but it was found that 
    one of the LeCroy 1151 modules does not clear at prestart.  However, it 
    does clear on start.  We are not sure why this happens for this module.  
    The other two 1151 modules that follow this one, clear in the prestart 
    stage.
    


    A copy of this log entry has been emailed to: camsonne,rbziel,melissac,tbadman,pzhu,rom,kelleher,slifer