    User name kamyers

    Log entry time 15:22:49 on March 08, 2012

    Entry number 364644

    keyword=[Target] red alarms on 4K supply temperature

    We are still getting the short excursions in the 4K supply temperature, which are apparently normal (related to something with filling of the dewer). Chris Keith changed the yellow alarm limit yesterday afternoon to get rid of these alarms, but I have gotten two red alarms recently as the last two excursions peaked slightly higher. The HIHI alarm limit is still set to 20K, while the HIGH limit is set to 30K. So, we might want to change that...



    A copy of this log entry has been emailed to: ckeith