    User name A. Camsonne

    Log entry time 13:13:49 on March 16, 2012

    Entry number 366690

    This entry is a followup to: 359397

    keyword=Check Right HRS magnet settings

    I compared Right arm setting to : http://hallaweb.jlab.org/halog/log/html/1203_archive/120308175005.html

    Right Arm Q1 momentum : 2.25115 Right Arm Q1 power supply : 1142.95 Right Arm Q1 gauss meter : -0.0285 Right Arm Q2 power supply current : 919.804 Right Arm Q2 momentum : 2.2513 Right Arm Q2 gauss meter : 0.4841 Right Arm Q2 power supply polarity : enabled Right Arm D1 power supply current : 788.247 Right Arm D1 momentum : 2.25098 Right Arm D1 NMR : 0.835311 Right Arm Q3 power supply current : 1084.01 Right Arm Q3 momentum : 2.25102 Right Arm Q3 gauss meter : 0.4805 Right Arm Q3 power supply polarity : disabled Right Arm B Field Hall Probe (D-G) : -0.8245

    This is pretty close to what we have, so no major magnet mistune and signs seem correct.