User name Jixie
Log entry time 22:46:45 on April13,2012
Entry number 374097
keyword=Septum current alarm is on
I have put a script to monitor the septa currents. It will give a voice alarm only if the following condition are satisfied:
A) C12 or NH3 target
B) Set currents is 0.5% deviated from the predicted value based on the HRS momentum and my "I vs P" curve; or read back currents are 0.5% away from the set points.
C) A and B last for more than 3 minutes.
This job is currently running on adaq@hapc7. Once alarmed it will sleep for 10 minutes.
If it is closed by accident, you can bring it back from this command:
On adaq@haapc7:
~adaq/beep/septaalarm