User name T. Badman
Log entry time 11:21:45 on April 22, 2012
Entry number 377451
This entry is a followup to: 377442
keyword=target amp fault
What is likely happening is when you move the target to one extreme (top
cell) and add such a large offset the encoder will think you are on the
wrong side of the limit switch. Trying to move away from that position
would cause a fault when it then tries to cross back over the limit.
When you get the amp fault is the target gui also saying you've hit the hi
or low limit? You can fix the issue rather easily in the hall by manually
jogging the motor back over the limit, then running home to reset the
encoder value, this might also work from PDP using relative jogging?