    User name A. Camsonne

    Log entry time 22:21:25 on May13,2012

    Entry number 384732

    Followups:

    keyword=BCM channel B stuck at 65535

    Channel B is stuck at 65535, I changed its calibration constant to 0.0000059 so it does not got into current calculation.

    for B :
    slope 0.0046875 offset -1

    New values for B : slope 0.0000059 and offset 0

    Channel A is unchanged and gives the right current, the current on the Hall A Wall is (Channel A + 0 )/2 so half of the current



    FIGURE 1