TJNAF Experiment 93-050
Charge Monitor Operating Directions
Table of Contents
General Overview
The Charge Monitor is composed of a Parametric Current Transformer
(also refered as Unser Monitor) surrounded by two resonnant cavities
(also called BCMs for Beam Current Monitors).
To run the control interface for the charge monitor:
- Logon on the hac computer (account hacuser, for password
ask the run coordinator) and type HAC_hp if you are on HP machine or
HAC_xt if you are on X terminal.
- Startup display will show up:
It summaries information about all the magnets, detectors and instruments.
To get to the Current Monitor screen, click on the BCM
button. The following screen will appear:
Calibration Instructions
- From the Hall A BCM screen, pull the Calibration screen:
- Call the accelerator people and organize the BCM calibration.
- Once they are ready to perform the calibration cycle, press
START button
- Once they are done with the cycle, press STOP button:
the two coefficients and their standard deviations should update.
Possible problems and Their Solutions
In case of any problem, check the Beam Current Monitor Calibration Manual,
or page Geraud at 680 6445.
Go back to the
E93050 Home Page
The content of this page is based on the similar text written by Segey Malov for E89033.
Please, send any comments on this page to
Geraud