2019/12/14 by Zhiwen Zhao Current boards are on floor. top row, adapter board "A021", it had maroc chips "726_728_729" bottom row, adapter board "A108", it had maroc chips "753_760_752" H13700-03 with SN PA0020 is mounted Two thing prvent a perfect mount 1. there is no gap between PMT and adapter board 2. distance between H13700 adaptor board and asic board is different from stack for H12700, I guess it means connector on H13700 adaptor board has different depth comparing to H12700 adaptor board So I have to unmount on rail, loose another rail, and use tape to mount it I mount PMT with a cut to match the HV cable. It seems correct But it's better to have a clear sign because connector seems go either way. Note only two chips on each asic boards are used "top 726_728_729" is placed on front side with PMT, 789 and 729 are used "bottom 753_760_752" is placed on back side without PMT, 760 and 752 are used Two asic board have opposite directions Somehow bottom LV cable is not working, so I use a desktop powersupply to keep it on, but it can't be controled remotely. (There are two desktop powersupplies at the bottom of the rack on right, I am using the left one) HV is provied by the SHV connnected cable in the black box. It's ch2 when you run hv script maroc HV cable are not connected LV and HV are not controled in the run script Do "lvon" and "hvon" before data taking Do "lvoff" and "hvoff" before data taking check "lvstatus" and "hvstatus" run script /home/rich/LaserTED/swFE/bin/RUNME_H13700.sh It will take top and bottom row data in two directories. make plot cd /home/rich/LaserTED/swFE/tile cd 726_728_729_2019_12_14_13_43 decodeFE run_000001.bin cd 753_760_752_2019_12_14_13_44 decodeFE run_000001.bin root 'plot_H13700.C+("726_728_729_2019_12_14_13_43/run_000001.bin.root","753_760_752_2019_12_14_13_44/run_000001.bin.root")' have fun!