Turn on power supply (contact James Proffitt, John LeRose,
Javier Gomez, or Mark Stevens for assistance).
If the NMR is locked (The NMR is locked if the field reading in
the blue field has a positive number and is updating itself). Don't
worry it will get there.
If the NMR is not locked (negative but updating number in the blue
field) but the current appears to be going in the right direction.
Don't worry it will get there and the NMR should lock when you get into
the vicinity of the desired field. If you're worried anyway and want to
watch the field change. Push SEARCH. (The NMR will search over the
full range of the probe rather than just in the vicinity of the desired
field.) The NMR should lock in a minute or two.
If the NMR is not locked and the current is going in the
wrong direction. (This can happen if you ask for a new field before you
reach a stable setting). You must push SEARCH. (The NMR will
search over the full range of the probe rather than just in the vicinity
of the desired field.) The NMR should lock in a minute or two and the
software will correct the current setting appropriately but slowly. If
the NMR doesn't lock you may have to ask for a field value appropriate
for the present current ( 1.1 10-3 T/Amp). The NMR
will ony search in the limited region of the requested field and should
find a lock more easily. After getting the lock let things settle in
for a few minutes and then ask for the field you want again. In extreme
cases you may have to nurse it through the transition by asking for
multiple small increments. If you ask for a change of less than 5% the
NMR should not lose its lock.
To shut down: Set FieldSet to .18. Let current run down to
just over 100 Amps. Then push stop on Dipole control screen
Resetting NMR Gaussmeter:
Problems
Explanation
Action
NMR not locked but current
Normal operation for large
Wait. (see
above)
is changing in the right direction.
field changes
direction
NMR locked but current going in
Normal operation.
Wait
going in the wrong
direction.
NMR locked but field not
Field regulation is disabled
Check
that field regulation
correct and current not
or software is confused.
is enabled. Enter
desired
changing
field value or one very near the
desired value again.
NMR field display freezes.
NMR Gaussmeter is not
Reset meter by
interrupting
(Usually but not always shows
communicating with
power
(see below)
-#.0000000)
software
All displays freeze
Computer is hung up
Reboot (see below)
Turn off power supply in rack in middle room ("Triple Output DC Power
Supply" second rack, facing counting room, head high with white sign
taped to front), wait about 20 seconds, turn back on. Press REM and
SEARCH on the control screen. You may have to repeat this a few times
to get it to work.
Rebooting: See Javier's written instructions (copied
below) or call either Javier Gomez or James Proffitt.
>rlogin hallasc16
>(user) hacuser
>(password) hacuser1
>ctr x (rebooting) (windows will go blank)
After rebooting:
Make sure field regulation is enabled then set the dipole field in the
dipole control window
set the current in the Q1 control window and still in the Q1
window make sure the following are set:
remote (comm)
unlock (fpanel)
negative (polarity)
Hit the OFF/RST button. (the box next to the "main power" will become
red)
Hit the ON button