Magnitudes of amplitudes from each of shower detector blocks has been equalized preliminary by the high voltage using signals from cosmic rays.
As shower detector is intended for electro-magnetic showers registration and identification, for the calibration constants definition it necessary to have events, where electron was registered by the shower detector.
Total absorption of electron energy is performed in Preshower and Shower, so for electron energy definition the energy deposition in both of these detectors must be summed up.
Calibration constants are fitted by minimization of functional
Here:
-- number of selected calibration event;
-- number of Preshower block, included into the cluster, reconstructed in
the -th event;
-- number of Shower block, included into the cluster, reconstructed in the
-th event;
-- set of Preshower blocks numbers, included into the cluster,
reconstructed in the -th event;
-- set of Shower blocks numbers, included into the cluster,
reconstructed in the -th event;
-- amplitude value into the -th Preshower block in the -th
event;
-- amplitude value into the -th Shower block in the -th event;
-- mean value of pedestal of the -th Preshower channel;
-- mean value of pedestal of the -th Shower channel;
-- particle momentum corrected for kinematic broadening in the
-th event;
, -- Preshower and Shower blocks calibration constants to be
fitted. Total number of these constants is 144 ( 48 of them for Preshower, and
96 for Shower).
Calibration events (i.e. events where the charged particle, detected in spectrometer, is electron with the biggest probability) are selected on the following conditions:
Calibration events selection is realized by program shcal_events, calibration constants calculation -- by program shcal_dofit (see sec.3.2).