New Moller Target Movement Control

Target Frame Parameters

  1. The targets are round foils with diameter 24.0mm see Picture 3.
  2. The distance between the foils centers of is: 1.5".

Linear Moving System

The stepper motor and the limit switches are connected to T-Channel of the motor controller VME44.
Basic Parameters:
  1. Number of positions : 6
  2. Range of motion from the "Retracted Limit Switch" to the "Extended Limit Switch" is: 356 mm
  3. Travel distance per one stepper motor revolution is: ~0.36 mm/revolution
  4. The number of steps per revolution is: 200
  5. In microstepping mode there is 10 microsteps/step
  6. The movement speed of the target frame is:~ 8-10 mm/sec

Linear Limit Switches (see Table 1):
  1. All the limits switches are "Normal Open" (gives logical(TTL) "1" to the controller) when they are not pressed. When the limit switch is pressed (a target in position or in one of limit position) it is closed and the logical "0" appears on the input of the controller.
  2. "Extended Limit Switch" is connected to the T-Channel (pin 23C of backplane VME44 connector)
  3. "Retracted Limit Switch" is connected to the T-Channel (pin 24A of backplane VME44 connector)
  4. All the targets switches are connected in parallel and look like one switch, which is connected to the "Home Limit Switch" of the T-Channel (pin 24C of backplane VME44 connector). The software should count the target switch crossings to cross check with the encoder reading for the target position.
To enable the stepper motors operations the "T Auxiliary Output"(22C) should be set to logical "0"

Signals to FSD System

  1. We suggest two modes of FSD operations:
    1. Normal mode an FSD signal is generated if the target frame is not in a "Target" or "Retracted Switch Limit" position.
    2. Moller mode The Moller FSD signal is not masked. This mode should be used only during Moller measurements. The target can be moved only with the beam turned off.

Software Requirements

  1. Initialization: It has to to be able to initialize system any time without iochla reboot
    1. Move the linear steppermotor to the "Retracted Limit Switch" of target frame and set this as the HOME position.
  2. Normal Operation Requirements:
    1. Move to Target 1-4 : a routine that moves the requested target on the beam.
    2. Move to Home Position ("Parking Limit Switch"): a routine that moves the target frame "Hole" on the beam.
  3. Software interlocks: there is an interlocks for movement of the target frame on or out of the beam.
  4. Only MCC can move target after the Hall A request.
  5. Off beam for any target movement.
  6. Beam is available only if the target position is on the "Parking Limit Switch" or on any "Target-N".



    Picture 1. View of Möller Polarimeter Target Assembly.

    Picture 1


    Picture 2. View of Target Loader Assembly.

    Picture 2


    Picture 3. View of Target Holder Assembly.

    Picture 3


    Table 1. Pinout OMS Chassis to Stepper Motor Controller VME44

    Signal Name P2 OMS Board P5 Connector Color OMS Chassis
    +5 Volts Channel T 21A b red +5V to linear MD125(10pin)
    T Step Output 21C c blue step pulses to linear MD125(9pin)
    T Direction Output 22A d yellow direction to linear MD125(8pin)
    T Auxiliary Output 22C e green "0"-enable Linear Motor
    Ground 23A f black Common
    T Extended Limit Switch 23C g orange Linear Right Limit Switch
    T Retracted Limit Switch 24A h brown Linear Left Limit Switch
    T Home Limit Switch 24C j white Linear Target On Beam Switch
     