Moller Measurements for E06-010 Transversity, Nov 2008 - Feb 2009
The Moller results are presented in a table below and on a plot 1
and plot 2.
# date day* Moller Moller with HALLA:p Syst.err. --------------------------------------------------------------------- 1 11.02.2008 12 +88.09 +/- 0.17 +88.09 +/- 0.17 +/- 1.7% 2 11.12.2008 22 -74.85 +/- 0.10 -79.47 +/- 0.10 +/- 2.0% 3 11.17.2008 27 -74.03 +/- 0.10 -79.63 +/- 0.10 +/- 2.0% 4 11.20.2008 30 -74.15 +/- 0.13 -76.94 +/- 0.13 +/- 2.0% 5 11.28.2008 38 +80.26 +/- 0.11 +78.03 +/- 0.11 +/- 2.0% 6 12.08.2008 48 -53.81 +/- 0.11 +54.10 +/- 0.11 +/- 2.0% 7 12.09.2008 49 -78.65 +/- 0.10 +79.08 +/- 0.10 +/- 2.0% 8 12.17.2008 57 -79.59 +/- 0.18 -78.47 +/- 0.16 +/- 2.0% 9 01.13.2009 63 -81.12 +/- 0.15 -79.88 +/- 0.16 +/- 2.0% 10 01.24.2009 74 -78.55 +/- 0.16 -79.88 +/- 0.16 +/- 2.0% 11 02.03.2009 85 -75.84 +/- 0.16 -79.14 +/- 0.16 +/- 2.0%* day - day of experiment running
Measurement 1: (11.02.2008)
There is no energy fluctuation. Beam energy correction was not done.
Measurements 2 - 11
The Moller measurements 2-8 were done at ~0.5uA curent. At this current HALLA:p energy reading is more noisy and for a part of
measurements energy correction can not be correctly done at all. This is the reason to increase systematic error from 1.7% to 2.0%.
Measurement 6: (08.12.2008):
From Eugine's words the beam polarization was low due to the injector problem.
Measurement 9: Spin Dance (01.13.2009)
The plot is given here .
The details of the measurements can be found here .