Moller Measurements for HAPPEX, July 1999


Last updated: Mon, 12 October 1999 moller@jlab.org

Here a summary is given of the Moller measurements done for the second period of HAPPEX running in 1999. The first period in described here.

During this period the laser phase was checked at every measurement.

The table of all the measurements is shown here:
   m   d   h   1-st run   Pol 1            Pol 2            Pol aver     chi2   uA   Las:power atten slit     loss  1/2wave B   C
   7  10  15    7006   75.70 +/- 0.20   75.40 +/- 0.20   75.60 +/- 0.20   1.6  0.502      73.0  500. 15.90    0.66   0      0   0
   7  12  17    7023  -74.45 +/- 0.18  -74.25 +/- 0.20  -74.35 +/- 0.18   0.5  0.540      73.0  500. 15.80    0.80   1      1   0
   7  12  18    7051  -74.85 +/- 0.20  -74.75 +/- 0.20  -74.80 +/- 0.20   1.8  0.140      73.0  500. 13.90    0.45   1      1   0
   7  13  17    7063  -75.05 +/- 0.18  -74.80 +/- 0.18  -74.90 +/- 0.18   1.5  0.510      73.0  500. 15.90    0.67   1      1   0
   7  15  14    7126   76.30 +/- 0.20   76.20 +/- 0.20   76.25 +/- 0.20   1.0  0.470      73.0  500. 15.50    0.98   0      0   0
   7  15  20    7146   76.50 +/- 0.20   76.30 +/- 0.20   76.40 +/- 0.20   1.4  0.550      73.0  500. 16.00    0.64   0      1   0
   7  15  21    7177   76.30 +/- 0.20   76.30 +/- 0.20   76.30 +/- 0.20   0.4  0.230      73.0  500. 14.50    0.52   0      1   0
   7  22  08    7189  -71.30 +/- 0.20  -71.10 +/- 0.20  -71.20 +/- 0.20   1.8  0.580      73.0  500. 15.75    0.92   1      1   1
where the date is given in the first three columns, the polarizations for 2 definitions of the target angle and their average are presented in the columns called "Pol 1" etc., and the additional attenuation parameter is shown in the column called "loss".

The Moller results, along with the Mott polarimeter results and indications of injector changes are presented on a plot.