Moeller Polarimeter: Results of the Runs with Unpolarized Beam, March 1998
Last updated: Tue, 24 Mar 1998
moller@jlab.org
Detector Tuning
ADC Pedestals
The pedestals were measured using pulser signals without beam.
ADC pedestals
ADC | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8
| 9 | 10 | 11 | 12 |
pedestal | 52. | 78. | 63. | 79. | 75. |
79. | 57. | 49. | 71. | 70. | 80. | 86. |
HV Tuning
High voltage of the two aperture counters was set to 1750V. For the Lead Glass
detectors ADC data were taken at various HV. The plots of amplitudes versus HV
are linear in the double logarithmic scale. After the initial adjustment
3 measurements were done at 4.045 GeV with 3 different high voltage values,
with a step of 50V for each channel. The left aperture (AL) counter was used for the trigger.
The run 1091 was taken with the optimized voltages which provided the Moeller
peak positions at the bin 300 in the ADC after pedestal subtraction, for
all LG modules.
HV for the test runs at 4.045 GeV
LG # | 1089 | 1090 | 1091 |
1 | 1730 | 1780 | 1680 |
2 | 1700 | 1750 | 1650 |
3 | 1715 | 1765 | 1665 |
4 | 1633 | 1683 | 1583 |
5 | 1664 | 1714 | 1614 |
6 | 1694 | 1744 | 1644 |
7 | 1731 | 1781 | 1681 |
8 | 1684 | 1734 | 1634 |
The Moeller peaks were fit and the dependence of the amplitudes on HV is shown
on a plot. Using these plots we calculated the voltages
which should move the peaks to the ADC bin 300 for other energies.
Such a setting is convenient since it allows the same discriminator level
independently on the hight of the hit in the lead glass column.
We should keep in mind that the dipole provides a dispersion of about
16% for the module size.
Calculated HV for various energies
LG # | 3.355 | 0.845 | 1.645 | 2.445 | 3.245 | 4.045 |
1 | 1696. | 1842. | 1770. | 1728. | 1699. | 1677. |
2 | 1664. | 1819. | 1743. | 1699. | 1668. | 1645. |
3 | 1680. | 1821. | 1751. | 1711. | 1683. | 1661. |
4 | 1599. | 1747. | 1674. | 1632. | 1603. | 1580. |
5 | 1631. | 1792. | 1712. | 1666. | 1634. | 1610. |
6 | 1660. | 1793. | 1727. | 1689. | 1663. | 1642. |
7 | 1700. | 1848. | 1775. | 1733. | 1703. | 1680. |
8 | 1652. | 1786. | 1720. | 1682. | 1655. | 1634. |
HV and Timing
The timing of the LG left and right sums slightly depend on the HV
with a slope of about -2 ns for 100V
(see a plot).
Energy Calibration
Runs 1227 and 1228 were used.
Events with a coincidence of all 4 signals in +/-10ns window were selected.
Calibration was performed in 2 steps:
- Calibration of the modules in the left and right columns
individually. Distributions of ADC-pedestal for all the modules
were filled. The peaks were fitted. The electron energy was assumed
to be E=0.5*(Y_center+Y_symm)/(Y_center+Y_module), where:
- Y_center=271cm*SIN(11o) - the distance from the beam
to the detector center.
- Y_module - the distance from the detector center to the center
of the module (=1.5*8cm for the upper modules).
- Y_symm - a shift of the symmetric pairs from the detector
center. This is the place where the electrons with an energy
of E0/2 would come. It was found iteratively to be about
-3.2cm (see the position measurements).
- It is assumed that the energy of an electron coming to
y=Y_center+Y_symm is 0.5 (of the beam energy).
Calibration factors CALIB were found such that CALIB*ADC=E.
- A fine tuning was performed. Histograms for the full energy
were filled for the cases when the maximum energy release happened
in the given modules of the left and right arms. Such a histogram
was fitted and the calibration factors of both modules
involved were moved in order to shift the full energy to 1.
Several iterations were done. This fine tuning did not change
the calibration factors but for 1-2%.
The histograms used for the fit are presented on
a plot where the top row
shows combination of Left 1 with Right 1-4, the second row shows
combinations of Left 2 ... and so on. The energy distributions in 2 arms
along with the total energy are also
presented.
Calibration factors 1./ADC_bin
ADC | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
calib | 0.002014 | 0.001571 | 0.001382 | 0.001180
| 0.001889 | 0.001601 | 0.001373 | 0.001243 |
The energy resolution is 5.5% at 4 GeV. Typical lead glass calorimeters
would provide something about 3-4%. In our case, however, a part
of the showers should escape to the sides making the resolution
worse.
Position Measurement
One can not expect a good position resolution because of a large
width of the lead glass modules and of considerable shower leakage to the
sides. Nevertheless, it should be possible to make a better estimate
of the hit position in the vertical projection than just the center
of the module with maximum energy.
The following procedure was used:
- Baricenter positions were found. Naturally, the baricenter distributions
have strong peaks at the centers of the modules
(see the plots).
It is also seen on this picture that the acceptances of two arms
are slightly different - it looks like the detector is tilted
in a way that the top is shifted to the left (looking downstream the beam).
- Assuming that the coming particles have a flat distribution along
the vertical axis Y (which is not really true) one can find a function
f(Yb) to correct the observed baricenter value Yb, such as
Y=f(Yb) by integrating the observed distribution for Yb.
It was done separately for the top (1,5) and the bottom modules (4,8)
with the distributions for L and R columns added, and for the modules
2,6 with the L ditribution added to an inverted R distribution
(in order to minimize the effect from a non-flat initial Y
distribution). The distributions obtained were integrated
and the hesulting histograms smoothed using his/oper/smooth in PAW.
The smoothed functions were used for f(Yb). The results are shown
on
(see the plots).
The resulting corrected distributions for the left and right arm
positions are better than the baricenters' distributions,
though display certain systematic shifts
(see the plots).
The two bottom plots present a position correlation of the hits
in both arms the left plot was obtained with no coincidence
condition while the right one - with 4-coincidence. A line on the right
plots shows the correlation expected from geometry of
the spectrometer.