we cleaned the dirty fiber head. We are using powermeter molectron powermax 5200 and probe molectron PM150-50 molectron was bought by coherent previously, the test was done a few mins after turn on laser. now we keep laser on for about an hour, and monitor the power output, it had no change. previously, the powermeter probe light in surface is about 20cm away from the laser fiber output. now it's about 10cm. we can't move it closer because the laser fiber is mounted on a optical holder and it can't go into the probe opening. This gives about 0.5W power increase comparing to previous measument. LD temp setting 25 when no light out LD temp 22+-1 Driver temp 23+-1 when light out LD temp 25+-2 Driver temp 51+-2 I/Imax% W(tested) W(provided) 50 6.7 7.6 80 18.9 20.8 95 24.5 27.2 There's still about 10% less than the manufacturer suggested