    User name vasulk

    Entry time Tue May 21 13:38:16 EDT 2002

    Entry number 327

    keyword=Analysis of 2nd Calibration

    Below is the analysis of the second water calibration with the gain at a factor of 100x and inverse on.

    Flux from the water cell: 42.83 cm^2
    Up signal (X and Y channels) divided by polarization: 1.040656e+09 +/- 2.086825e+07
    Down signal (both channels) divided by polarization: 1.000053e+09 +/- 2.255035e+06
    Total signal (up and down both channels): 1.020354e+09 +/- 1.536231e+07
    Pre-Amp amplification: 100x
    10k-100k Filters on

    Water calibration constant: 0.07310 +/- 0.00189 mV/(%*amagat*cm^2)

    Note: This calibration agrees within error bars compared to the first calibration at 100x gain and inverse off.