Beam startup/recovery Procedure:
This procedure is required for major beam down: more than 2 hours or so.
- Initial requirements:
- Beam energy lock: ON Hall A
- Fast feedback lock: ON
- Target: Empty target
- Raster: OFF
- Setup beam to Hall:
- Let MCC tune beam through Compton chicane with 2 μA beam.
- Make sure BPM A and B readings are within accepted range (see whiteboard).
- Harp Scan:
- Ask MCC to do a harp scan, make sure beam σx,y within 100 - 300 μm on both hall A harps.
- Beam position check with BeO target:
- Move target to BeO, ask for 2μA beam with raster off.
- Make sure the beam spot on the camera matching the marker.
- Ask for 4x4 (MCC unit) raster, take L-HRS data with T1 trigger only. Do spot++ to verify the raster size: 3x3 mm2.
- Clearance check on LD2 target:
- Move to LD2 target, keep 4x4 (MCC units) raster and 2 μA beam.
- Check rate with reference number posted on the right side of the scaler monitor.
- Take L-HRS T1 data and check raster size with spot++: 3x3 mm2. Make sure no sign of cell wall hitting.
- Ask for 6x6 (MCC unit) raster and make sure rates stay the same.
- One more run and spot++ to check raster size, 4x4 mm2, and no touching of cell walls.
- Now beam is good to go:
- Make sure the VDC high voltage is OFF for production or any high rate tests.
- Ramp current slowly, keep an eye on the LD2 target temperature.
- Start to take data and have fun.