• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name S. Sirca

    Log entry time 06:22:35 on December11,2000

    Entry number 53319

    keyword=BPMs A/B x/y do not update

    Whatever the beam condition, BPM B constantly shows x=-0.110, y=-0.084
    on both HAC and GEN_O (from MCC) displays. MCC induced a small kick
    at the end of the beamline to see if the BPMs respond, but they do not.
    They reset the corresponding IOC and we reset ROC14 (same guy?),
    but it only caused CODA not ending a run properly, and the x,y values were
    still the same on HAC. However, the numbers on GEN_O are now updated.