    User name adev

    Log entry time 06:22:38 on December 17, 2000

    Entry number 53715

    keyword=End_of_Run_1954
    ****************************
     Run Number: 1954
    End of Run Comment:
    Tar=LH2-15cm,Type=Elastic,Lmom=3.3593,Rmom=1.0,Ene=4.58,Angle_R=12.509,Angle_L=23.768
    
    
    Time of run = 6.9727213 min
    Av. current (on 3x) = 15.359857 uA
    Av. current (on 1x) = 15.321851 uA
                   |   on 1x    |   on 3x    |   on 10x
    Av. charge (mC)|    6.4101  |    6.4260  |    6.3802
    up-down    (mC)|    0.1030  |    0.0662  |    0.1184
     Trigger   | Scaler/charge(3x)|Scaler/time/ps|  Scaler    | Prescale| Scaler/charge(1x)
        1      |   6428919.7 /mC  |    0.549 Hz  |  41312238  | 180000   |   6444866.4 /mC
        2      |    102414.4 /mC  |    0.315 Hz  |    658115  |  5000   |    102668.4 /mC
        3      |      7167.1 /mC  |  110.086 Hz  |     46056  |     1   |      7184.9 /mC
        4      |    799380.6 /mC  |    6.139 Hz  |   5136820  |  2000   |    801363.5 /mC
        5      |      1021.5 /mC  |    0.000 Hz  |      6564  | 65535   |      1024.0 /mC
        output                        35.431 Hz
    Av. deadtime from accepted triggers = 0.6974
    Av. deadtime from gated scalers = 1.0000
     Trigger   |  Plus Helicity   |  Minus Helicity |  Sum 
        1      |  98679.884  Hz     |    98822.815 Hz   |  98747.285 Hz
        2      |  1569.5067  Hz     |    1576.7604 Hz   |  1573.0707 Hz
        3      |  110.35942  Hz     |    109.86628 Hz   |  110.08614 Hz
        4      |  12280.172  Hz     |    12279.788 Hz   |  12278.372 Hz
        5      |  15.578446  Hz     |    15.814709 Hz   |  15.689713 Hz
        output |  0.0  Hz   |    0.0 Hz |    35.431 Hz
    Accepted triggers  =  14823.00000
     Time of run = 6.9727213 min
     Time of run h- = 3.4461588 min
     Time of run h+ = 3.4470703 min
     1-(sum of time hel + and -)/( run time) = 1.1400e-02
    Rate of Left Arm Pulser for electronic DT  = 10.156245
    Rate of Right Arm Pulser for electronic DT  = 5.0028291
    Rate of Left Arm Pulser&trig for electronic DT  = 10.151465
    Rate of Right Arm Pulser&trig for electronic DT  = 5.0028291
    Left Arm Electronic DT  = 0.00047069899
    Right Arm Electronic DT  = 0.0
    Rate of Left Strobe  = 21214.854
    Rate of Right Arm Strobe  = 0.0
       helicity    |  Gate on + and run  |  Gate on - and run  |  Gate on run 
          +        |  11.246207  Hz         |    0.0 Hz        |  11.246207 Hz
          -        |  0.0  Hz        |    11.243817 Hz       |  11.246207 Hz
     
        Signal     |  Gate on + and run  |  Gate on - and run  |  Gate on run  |
                   |     Charge   (C)    |     Charge   (C)    | Charge   (C)  
       upstream    |   0.0031929         |   0.0031956         | 0.0064616     
       upstream 3x |   0.0031917         |   0.0031944         | 0.0064591     
       upstream 10x|   0.0031821         |   0.0031845         | 0.0064394     
      downstream   |   0.0031419         |   0.0031447         | 0.0063586    
     downstream 3x |   0.0031590         |   0.0031617         | 0.0063929    
     downstream 10x|   0.0031236         |   0.0031260         | 0.0063210    
     
     
        Signal     |  Asymmetry
                   |           
       upstream    |   4.22634420e-04 
       upstream 3x |   4.22793250e-04 
       upstream 10x|   3.76967300e-04 
      downstream   |   4.45391790e-04 
     downstream 3x |   4.27167880e-04 
     downstream 10x|   3.84024580e-04 
     
    
    ****************************
    End-of-run EPICS data for Run Number 1954 at Sun Dec 17 06:22:24 EST 2000
    Right Arm D1 momentum                    : 1
    Right Arm arm angle                      : 12.509
    Left Arm D1 momentum                      : 3.3593
    Left Arm arm angle                        : 23.7679
    Tiefenbach Hall A energy                : 4582.78
    Right Arm arm mispointing (HPT)          : -1.17921
    Left Arm arm mispointing (HPT)            : 5.29134
    Right Arm Q1 power supply current        : 752.01
    Right Arm Q1 momentum                    : 1.0001
    Right Arm Q1 gauss meter                 : -0.018772
    Right Arm Q1 power supply polarity       : 0
    Right Arm Q2 power supply current        : 430.156
    Right Arm Q2 momentum                    : 1
    Right Arm Q2 gauss meter                 : 0.22549
    Right Arm Q2 power supply polarity       : 0
    Right Arm D1 power supply current        : 342.996
    Right Arm D1 momentum                    : 1
    Right Arm D1 NMR                         : 0.371088
    Right Arm Q3 power supply current        : 398.16
    Right Arm Q3 momentum                    : 1.00063
    Right Arm Q3 gauss meter                 : 0.17789
    Right Arm Q3 power supply polarity       : 1
    Left Arm Q1 power supply current          : 2527.59
    Left Arm Q1 momentum                      : 3.35968
    Left Arm Q1 gauss meter                   : 0.21418
    Left Arm Q1 power supply polarity         : 1
    Left Arm Q2 power supply current          : 1446.52
    Left Arm Q2 momentum                      : 3.35931
    Left Arm Q2 gauss meter                   : -0.7613
    Left Arm Q2 power supply polarity         : 0
    Left Arm D1 power supply current          : 1192.66
    Left Arm D1 momentum                      : 3.3593
    Left Arm D1 NMR                           : 1.24416
    Left Arm Q3 power supply current          : 1337.35
    Left Arm Q3 momentum                      : 3.35918
    Left Arm Q3 gauss meter                   : 0.6672
    Left Arm Q3 power supply polarity         : 0
    Average current of the two cavities     : 0.0316001
    Current on Unser monitor                : -2.48164
    Current on Upstream bcm                 : 0.0293172
    Current on Downstream bcm               : 0.033883
    Upstream Cavity Coefficient             : 77
    Downstream Cavity Coefficient           : 78.9
     Current on IBC0L02 (source)            : 0.00142881
    MBSY1c Hall A Beam energy               : 4592.25
    MBSY1c magnet current (amp)             : 191.553
    MBSY1c Bdl                              : 1.146e+06
    MBSY3c Hall C Beam energy               : 2329.44
    Tiefenbach Hall A energy                : 4582.78
    Tiefenbach Hall C energy                : 2323.99
    Delta(E)/E width (MV_AI_DXWID)          : 2.12374
    IPM1C12.XPOS                            : 0
    IPM1H04A.XPOS                           : 0
    IPM1H04A.YPOS                           : 0
    IPM1H04B.XPOS                           : 0
    IPM1H04B.YPOS                           : 0
    Right Arm Arm Collimator                 : 0.363965
    Left Arm Arm Collimator                   : 0.868506
     HacR_HPM_LVDT:VALr                     : -1379
     HacR_VPM_LVDT:VALr                     : -1663
     HacR_HPE1313A:0_Ch58                   : -0.569824
     HacR_HPE1313A:0_Ch55                   : 22.7051
     HacR_HPE1313A:0_Ch54                   : -0.84375
     HacR_alignROLL                         : -0.363132
     HacR_alignPITCH                        : 0.0116197
     HacR_alignHPT                          : -1.17886
     HacL_HPM_LVDT:VALr                     : -8111
     HacL_VPM_LVDT:VALr                     : 1371
     HacL_HPE1313A:0_Ch58                   : -0.129395
     HacL_HPE1313A:0_Ch55                   : 23.7305
     HacL_HPE1313A:0_Ch54                   : -0.281738
     HacL_alignROLL                         : 0.0821989
     HacL_alignPITCH                        : -0.129473
     HacL_alignHPT                          : 5.29152
    haBDSPOS                                 : 5.354e+06
    haL1_Fan_frequency_R                     : 0
    haL2_Fan_frequency_R                     : 62
    haL3_Fan_frequency_R                     : 0
    haH2_P_Fill_Target_R                     : 23.12
    haH2_P_Exhaust_R                         : 23.28
    haH2_P_Fill_Bottle_R                     : 23.56
    haH2_DP_Fill_Exhaust_R                   : 0
    haD2_P_Fill_Target_R                     : 0
    haD2_P_Exhaust_R                         : 0
    haD2_P_Fill_Bottle_R                     : 0
    haD2_DP_Fill_Exhaust_R                   : 27.686
    haSC200_6_Sensor_1_R                     : 242.45
    haSC200_6_Sensor_2_R                     : 17.15
    haSC200_6_Sensor_3_R                     : 3.07
    haSC200_6_Sensor_4_R                     : 40.18
    haHe_Omega_DPG_Sensor_R                  : -0.03
    haITC502_1_Sensor_1_R                    : 94.98
    haITC502_1_Sensor_2_R                    : 90.19
    haITC502_1_Sensor_3_R                    : 94.98
    haITC502_2_Sensor_1_R                    : 19
    haITC502_2_Sensor_2_R                    : 19
    haITC502_2_Sensor_3_R                    : 19.03
    haITC502_4_Sensor_1_R                    : 76.6
    haITC502_4_Sensor_2_R                    : 19
    haITC502_4_Sensor_3_R                    : 67
    haITC502_5_Sensor_1_R                    : 62.39
    haITC502_5_Sensor_2_R                    : 60.74
    haITC502_5_Sensor_3_R                    : 154.3
    haITC502_6_Sensor_1_R                    : 96.75
    haITC502_6_Sensor_2_R                    : 99.97
    haITC502_6_Sensor_3_R                    : 26.3
    Right Arm Arm Collimator                 : 0.363916
    Left Arm Arm Collimator                   : 0.868408
    Wien angle                               : -6.99926
    Helicity Mode ON/OFF Random/Toggle       : OFF
    Wien BDL                                 : -203.2
    Wien field                               : 5.6245
    Laser attenuation Hall A                 : 120
    Laser power  Hall A                      : 73
    Half-wave plate (off=out)                : ON