Log entry time 21:19:11 on March15,2001
Entry number 58853
keyword=Warning
Incident with Q2 Left teaches us:
it is not enough to write down gaussmeter/current numbers on the run sheet.
it is very important to actually check that the numbers are correct.
One cannot rely on the momentum flaging red. It does not if the DVM freezes,
and regulation see a frozen (and tehrefore "correct"curent). The field will drift, the momentum number will not change.
Once more: whenever numbers are written down, they must be compared with the reference list which will be posted next to the magnet page monitor from now on.