User name X. Zheng
Log entry time 04:51:48 on July 2,2001
Entry number 65382
keyword=Target - EPR - 38.2%
4:31am: EPR during data taking, beam on
gradient coil current L=-2.00A, T=5.00A
Polarization:
1st order fit:
down: freq. shift = -46.08 KHz +/- 3.79 KHz, P = -38.27%+/- 3.15%
up: freq. shift = 45.65 KHz +/- 4.04 KHz, P = 37.92%+/- 3.35%
0th order fit:
down: freq. shift = -45.89 KHz +/- 3.84 KHz, P = -38.12%+/- 3.19%
up: freq. shift = 46.08 KHz +/- 4.40 KHz, P = 38.28%+/- 3.66%
note: I furthur decreased rel. gain by about 0.4 turn to reduce the noise. So during
the down flip I have to turn off integration. The circuit also cannot track the
resonance during the up sweep. It finally went back to the original freq when I put
modulation source to be 2.0 Vpp. For analysis we have something to fit, (and it is
important to know where the system starts to sweep, where it starts to lock in, in one word, which point should we fit). but the spectrun looks really bad.
Suggest next time use high mod. source amplitude (2.0Vpp?), but I am not sure. The system cannot track sweeps with 0.7Vpp.