    User name J.P. Chen, W. Korsch

    Log entry time 10:52:31 on July25,2001

    Entry number 67513

    keyword=Target - NMR: 31.3%, EPR: 33.1%

    10:10 am NMR/EPR measurements

    Conditions: beam off, target position: pick-up coils, all grad. coils off, B-field: 0 deg.,
    all lasers off (for NMR), last run numbers: 2250/21250

    NMR: P = 31.3%

    10:19 am lasers# 5,6,7 on, B-field: 0 deg
    10:26 am temperatures stable

    EPR: P = 33.1%

    1st order fit:
    freq1: -35.97 kHz +/- 1.75 kHz | Polarisation: 32.84 % +/- 1.59 %
    freq2: 36.52 kHz +/- 1.49 kHz | Polarisation: 33.35 % +/- 1.36 %

    Comments: good sweep, V_mod=0.8V, V_RF(EPR) = 16.0V (Wavetek),
    B-field was still drifting during sweep, so performed 1st order fit only.