Main INDEX
Monthly INDEX
PREV
NEXT
User name weinstein
Log entry time 19:03:22 on November29,2001
Entry number 78304
This entry is a followup to: 78302
Followups:
keyword=beam spot size
MCC did a harp scan to double check the OTR results (see halog 78302)
It turns out that the OTR calibration is too big by a about a factor of 5.
The actual beam is sigma_x = 60 um and sigma_y = 360 um (see the accelerator log for details) Note that these are accelerator coordinates so y is vertical (and corresponds to our x_target)