    User name fissum

    Log entry time 16:53:46 on December 6,2001

    Entry number 78825

    keyword=guards left in place - Emiss problem solved

    We found the solution to our Emiss shift for the high Emiss G data.

    Unfortunately, the target window shield and caps for the HRS were
    left in place - we believe from maintenance Tuesday 12/04. This
    amounts to a little plastic and a few millimeters of aluminum in the
    way of both the scattered electron and knocked-out proton. Fortunately,
    we think we can use all the data we took. We are pretty pleased that
    we caught this in the data so quickly.

    This was due to a little bad luck - the spectrometer configuration
    was such that the caps on the HRS were not visible given the
    camera locations. Further, those on shift were not sufficiently
    familiar with the way the plastic guards looked to be able to
    recognize they were in place.

    Recommendations:

    1. another camera in the hall mounted in a very different location
    from the two we presently have.

    2. the clear plastic windows used to guard the scattering chamber
    window should be painted yellow or completely covered with the
    black/yellow striped tape so that it is a little more recognisable that
    there is something on the target.