    User name R. Michaels

    Log entry time 01:22:30 on May23,2002

    Entry number 4030

    keyword=raster and intrinsic spot size

    We spent about an hour trying to get the rastered spot to approx 2 x 2 mm.
    Finally after tuning quads we achieved this. Then we did a HARP scan to
    check intrinsic spot size. Harp run 1722, CODA run 1802. Result was
    61 x 99 um at BPM4A and 144 x 27 um at BPM4B. We ask MCC to adjust
    Y size larger. They are trying...

    Fig 1 are fits to our event BPM data. The sigmas are 72 x 81 um at BPM4A
    and 92 x 95 um at BPM4B but the BPM resolution is probably 20 sqrt(100/I)
    microns where I = beam current in microamps (I think).


    FIGURE 1