• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name R. Michaels

    Log entry time 03:40:01 on June16,2002

    Entry number 86126

    keyword=R-HRS Plane 1 Pad 1 PMT-R missing in scaler

    On the right HRS, plane 1 scintillator, paddle 1, right-PMT the scaler data
    is zero. But this PMT must be working, according to other info (e.g. the "and"
    of L+R PMT is nonzero rate, and the data shows up in ADC's and TDC's).
    This channel of scaler or something leading up to it is not working; perhaps
    one of the ribbon cables is pulling out of the sockets.