    User name Higinbotham

    Log entry time 16:15:28 on January17,2003

    Entry number 92276

    keyword=Beam Spot Size Checks

    Looking at the raw Harp data, our beam size is clearly larger than the
    finally numbers are indicating.

    The raw width of the beam given by three harp wires is 201,171, and 183 um.

    It looks like the problem is the code is trying fit an ellipse, but as we can see on the OTR, our present beam is not elliptical...