• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name Armstrong

    Log entry time 09:42:56 on July10,2004

    Entry number 128104

    keyword=When the septa temp readout died

    
     Looking through the septa temperature history (using the
    JavaArchiverClient, which is available from the Hall A tool
    button on the hacuser desktop, and which is fairly easy
    to use), I see that the 6 thermistor values from the two
    septa (upper coil, lower coil, yoke), all flatlined
    at the same time, and that was July 6, just after 10:00 AM.
    Don't know if this information is useful,,,
    
    
    One lesson: if the septa strip charts don't show that these
    temperatures vary (with beam off) at the least significant
    bit (0.05 K) level, then it means that they are not being
    readout properly. Another clue is if beam-off temperatures
    don't match well with previous data - eg, if the right
    septum upper coil is not the highest of the temps, there
    is likely a readout problem. 
    
    The figure below shows the temperature history from
    July until July 8; one sees beam on/off data at the
    beginning, lots of beam-off data with valid readout, then
    the flatline traces.
    


    FIGURE 1