User name gilman
Log entry time 16:07:35 on July18,2004
Entry number 128872
keyword=day shift summary
Run 2939 running as we come in - note typo in owl shift summary.
Plan for day: production data. Problem for now is that the
Hall C charge asymmetry is large, ~100 ppm, rather than < 10 ppm
as it should be.
08:14 run 2940 started
09:11 run 2941 started
10:09 run 2942 started
The Compton DAQ has crashed again. I restart it. The auto-
analysis of the last run appeared to fail. In the new run,
it appears to only rarely lock, so I guess we are mostly
measuring background.
11:10: 0.52 % bleedthrough measurment, half wave plate inserted
11:10 run 2943 started
12:06 run 2944 started
cryotarget ioc rebooted at ~12:30
13:03 run 2945 started
14:00 run 2946 started - flipper inadvertantly run just before run started
***septum lost - right septum tripped***
Trip occurred about 1:00, no alarm noticed. No big jumps in
beam current. Lumi problem Astrid noticed may have been partly
due to this.
while recovering septum, rfzone problem and no bea, for rest of shift.
kent ans lisa take over to fix up charge asymmetry and recover septum
during this time.