Main INDEX
Monthly INDEX
PREV
NEXT
User name Geraud
Log entry time 10:34:52 on October16,2004
Entry number 134393
keyword=bogie reboot
We had trouble when trying to move the LEFT HRS. The bogie was frozen and would not work. We tried to kill it but it wouldn't restart, even from downstairs.
Then we tried to close all the MEDM windows but that didn't help.
Then we called Heidi that suggested the power of the spectrometer movement had to be turned OFF and ON from downstairs (back of spectro) but that didn't help either.
We found an old entry suggesting that a crate might need to be rebooted, but we didn't manage to do it from the Hall A general tools.
Mark Stevens called also. The final trick was to use the green button from the counting middle room to reboot the Left spectrometer bogie IOC. It works now.