• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name Carlos

    Log entry time 19:41:15 on November 8,2004

    Entry number 136183

    keyword=TS1 not responding


    We had a couple of times this error from TS1, and DAQ crashed.
    A single TS1 reboot and restart coda is not enough.
    Kill coda, reboot all TS1, all rocs and start coda again works.

    informEB: msgQSend done...
    prestarted
    daLogMsg: prestarted
    activating
    daLogMsg: activating
    informEB: msgQSend done...
    rolp->daproc = 5
    Entering User Go 2
    daLogMsg: Entering User Go 2
    rolp->daproc = 5
    Entering User Go
    daLogMsg: Entering User Go
    active, events so far 1 token 0
    daLogMsg: active, events so far 1 token 0
    nevents 1 newevnb 255 old 0
    nevents 1 newevnb 257 old 255
    nevents 2 newevnb 255 old 1
    nevents 2 newevnb 258 old 255
    rolp->daproc = 3
    VME Interrupt mask=0x0 is out of range
    User End Executed
    daLogMsg: User End Executed
    rolp->daproc = 3
    User End 2 Executed
    daLogMsg: User End 2 Executed
    ERROR; write_thread failed (in recv).
    rolp->daproc = 3