• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name R. Michaels

    Log entry time 09:29:07 on August 9,2005

    Entry number 151245

    Followups:

    keyword=cavities du jour (current scan)

    Last night there was a test of the beam current ramp, run 4224.
    Here are some cavity monitor results. Cavities allegedly work
    since Thurs Aug 4 at 17:00, so let's confront this claim.

    Fig 1
    BCMCAV2 and BCMCAV3 vs BCM1 -- looks pretty good.
    Some evidence of a nonlinearity at low currents, but this is expected.
    Freyberger says the cavities need automatic gain adjustment. His group
    will implement this, but in the meantime the settings are only good
    around our operating current (ca 35 uA).

    Fig 2. Position correlations during the same run.
    Cuts on bcm1>8000 to ensure the beam is high (reason: see above).
    If I don't make this cut the results are not as pretty.

    A) BPMCAV2X vs BPM4AX looks well correlated, but...
    B) BPMCAV2X vx BPM4BX is not well correlated. I'm not as worried
    about this now, see comments below. BTW, this implies that BPM4BX
    is not well correlated to BPM4AX (stripline to stripline) and
    nobody is worried about that, right ?
    C) BPMCAV2Y vs BPM4AY looks sort-of correlated.
    D) BPMCAV2Y vs BPM4BY is more correlated than plot C).
    The above 4 plots were concentrating on CAV2 positions.
    Now, running out of room in my plot, I plot the 2 prettiest
    correlations for CAV3:
    E) BPMCAV3X vs BPM4AX
    F) BPMCAV3Y vs BPM4BY

    Discussion: Between each monitor the beam is changing a lot
    (need for a big phase advance). What's more, the beam is moving
    rapidly, and there are electronic phases between devices which are
    a significant fraction of the raster period. Finally, each device
    has its own integration time. I think that combining these effects
    cause the imperfect correlations we see, and it is not a problem.
    To test this, we could do a "bulls eye" scan, where the raster is
    nominally off and the beam is moved slowly and parked in different
    spots. Regarding the need to have raster-off data, my understanding
    is that this would still be useful to fine tune the phase zero
    offset, but the present value used should be pretty close.

    As for the info I had Friday about phases being unstable, this was
    explained in
    halog 150700.

    According to Musson this does not affect our measurements.
    It makes me worried but they say we should be ok. We'll see.



    FIGURE 1

    FIGURE 2