    keyword=Cavity update / summary of status

    Cavities look fine this evening, after Musson's gain change earlier
    today (halog 152004). Spectr. Run 2468. See fig 1 & 2, Amplitude distribution
    and correlation to striplines looks good but not better than before.

    Summary of the status:
    1. The noise floor causes the glitch in the amplitude distribution
    in the middle of the raster orbit. We may eventually fix this by increasing
    pre-gain but the implications for HAPPEX are negligible if
    the resolution is as good as we think.
    2. HAPPEX DAQ resolution is <= 10 um. Seeing if it is better than
    this requires a careful regression analysis. Spectrometer DAQ
    resolution appears to be worse (expected), prelim: <= 120 um.
    3. The phase offsets and gains are fine. We don't need to touch
    them anymore during HAPPEX. The insensitivity to phase offset is
    understood but I won't try to explain it here.
    4. The frequency response has a bandpass of 60 kHz (approx). This
    does not matter for HAPPEX, and it may eventually be improved
    to 200 kHz, which would be helpful to counting experiments.
    5. There have been rare reliability problems. The PLL may get out
    of lock, requiring an IOC reboot. When the reboot occurs the default
    gains may be wrong (this will be fixed ASAP, I'm told). We must
    improve the routine diagnostics and recovery procedures.
    6. Diagnostic: Run Podd analyzer on a 50K run, as you would for
    spot++, and run the macro as directed:   
    .x online.C+("cavities")