    User name sarty

    Log entry time 21:58:15 on July22,2006

    Entry number 174859

    Followups:

    keyword=Harp Scan (wide beam)

    Did a harp scan. Looks like the vertical wire in Scanner 5 is broken (never a signal from the right-most "peak" in scanner 5). Scanner 6 looks ok. Harp data file = scan_2216.data
    Size of beam from Scanner 6 is: horizonal sigma = 419 microns, vertical sigma = 543 microns.

    Will do an OTR check now to see if this is consistent with that.