Main INDEX
Monthly INDEX
PREV
NEXT
User name jpchen
Log entry time 08:03:45 on November 9,2007
Entry number 213160
keyword=owl shift summary
Owl shift summary.
Shift workers: J. P. Chen (SL), Y. Zhang (TO), R. de Leo (3rd).
0:00 Finished raster set-up: MCC 5x3 gives us 2x2.5 mm.
Started beam x scan with empty target:
0, 1, 2, 3, then -1. -2, -3, -4, -5.
Found x is offset by 1 mm. Will ask for beam to be (-1,0) in both BPMA and BPMB.
1:00 Spot_L shows a problem: no BPMA, BPMB strange shape.
1:10 Moved Pb target in.
Beam x scan -1, -2, -3, 0, 1, 2
Did -1,-2,-3 before controlled access.
2:23 requested controlled access to let Alexandre to
look into the left BMP (SOPT_L) problem.
changed left momentum to 215.105 MeV to have higher rate for
NaI calibration.
3:50 ended controlled access.
3:55 beam back. resume x scan on Pb (0,0)
4:15 (1,0)
4:25 (2,0)
4:30 requested beam to be (-1,0), and then increase to 50 uA for
production.
4:38 started production with 50 uA.
4:40 right dipole quenched.
4:50 Alexandre changed NaI HV.
4:50 reduced beam current to 40 uA, changed PS4 to 10.
4:55 beam off, vacuum valve problem.
5:45 changed left P back to the first setting, 813.763 MeV.
moved target to He4.
6:47 beam back.
7:30 beam off, ioc reboot.
7:56 beam back
Left spectrometer, 120 degree.
p=215.105 MeV
2115, 4:38-4:42, 271K, 50uA on Pb, PS3=PS4=1, deadtime 20%.
2116, 4:53-5:03, 97K, 40uA on Pb, PS4=10, deadtime 13%.
p=813.763 MeV.
2117, 6:47-7:02, 222K, 50 uA on He4.
2118, 7:15-7:22, 225K, 50 uA on Pb
2119, 7:56-8:02, 112K, 50 uA on 10 cm dummy.
Right spectrometer, 120 degree.
p=534.589 MeV
20622, 4:38-4:42, 132K. 50 uA on Pb, right dipole quenched during the run.
20623, 7:15-7:21, 247K, 50 uA on Pb, dipole is 0.2% off at the begining, settled to 0.1% during the run.
20624, 7:56-8:02, 117K, 50 uA on 10 cm dummy.