User name Sirish
Log entry time 17:30:04 on November13,2007
Entry number 214337
Followups:
keyword=Compton Back Online
Compton photon DAQ problem is fixed. Replacing a cable on the trigger line seems to do the trick.
For running the background check the guidelines are as follows:
1. Photon detector HV=-1500 V (Beamline crate, slot 11 chan 1)
threshold=10 mv
---- DO NOT CHANGE THESE VALUES ----
2. < 10kHz/uA in photon counting rate is good beam
3. If rate exceed 10 kHz in a sustained manner action has to be taken.
a) if rates are 10 - 100 kHz request MCC to reduce compton photon
rate, ususally via "dogleg pathlength" corrections. This is
generally non-intrusive to the Halls.
b) if rates are > 100 kHz, the injector tune is likely to be bad.
Request MCC to check injector. This is intrusive and takes a
while.
c) if rate > 1 MHz turn off photon detector HV.
4. After every machine retune, e.g. Energy Change, First go straight
bypassing Compton Chicane. Check for momentum spread < 5E-5 in the
SLI and visually check for a clean symmetric beam on the target
OTR. If these check out ok, turn on the Chicane to monitor photon
rate.