    User name V. Sulkosky

    Log entry time 19:01:19 on November13,2007

    Entry number 214338

    keyword=LHRS BPM Problematic History

    Listed below are the periods of time that BPMA and BPMB read-out for
    the LHRS was impaired by one or more ways.


    Pre-Access on November 7, 2007:
    - Runs before and including 2088 are fine.

    Controlled Access on Nov. 7, 2007 (Day shift):
    - Run numbers 2089 to 2110 BPMA and BPMB Antenna # 3 cables were
    unplugged to the LHRS.

    Controlled Access on Nov. 9, 2007 (Owl shift):
    - Run numbers 2111 to 2123 BPMA and BPMB Antenna # 3 cable
    were still unplugged to the LHRS.

    - Alexandre added a new attenuator box to the LHRS crate and
    connected the signal cables for BPMA Antenna #1, 2, and 3 to
    this new module.

    - During this period, the BPMA antenna #2 signal was saturated.

    - The signal from BPMA antenna #1 is also affected, since the
    attenuation and offset were not adjusted.

    Controlled Access on Nov. 9, 2007 (Day shift):
    - Run numbers 2124 to 2128 were cosmics tests for the BPM signals.

    - BPMA and BPMB Antenna #3 cables were reconnected.

    - The attenuation for BPMA antennas 2 and 3 were adjusted to match
    the outputs from the other antenuator modules, 0.08 volts.

    - The offsets were not adjusted for antennas 1, 2 or 3 (BPMA).

    - Run numbers 2129 to 2329 have an offset problem for BPMA.

    Controlled Access on Nov. 12, 2007 (Swing shift):
    - All run numbers starting from 2330.

    - The BPMA antennas 1, 2 and 3 are reconnected to the original
    attenuator box with correct settings for the attenuation and
    offsets.