Main INDEX
Monthly INDEX
PREV
NEXT
User name V. Sulkosky
Log entry time 15:37:42 on November21,2007
Entry number 215979
keyword=BPM positions after Ped Update
Figure 1 shows the BPMA and BPMB positions after I updated the
pedestals in the db_rb.Raster.dat, db_urb.BPMA.dat and db_urb.BPMB.dat
database files for the CSR database directory: DB/20071108. Figure
2 shows the BPM positions before updating the pedestals. The plots
are for Run 2150, See Halog entry 213381
Unfortunately, updating the pedestals made matters worse. Hence, the
calibration constants are incorrect with data using the "new"
attenuator module.
To fix this, I will first calibrate the BPMs from the Bullseye scan
using the original attenuator. I might then be able to use the
RHRS BPM information to extract a calibration during the period of
using the "new" attenuator on the LHRS.
FIGURE 1
FIGURE 2