• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name V. Sulkosky

    Log entry time 17:26:10 on January 18, 2008

    Entry number 226318

    This entry is a followup to: 226270

    keyword=Cavity Monitor in RHRS Update

    The past two days I investigated the problems that I saw with the cavity monitors. Except for one of the signals, downstream y position, the other signals appeared to be saturating at low ADC channels. After further study, I believe I had attenuated the signals too much.

    1) Today I used a 1.5 V battery, which is comparable to the signal expected from the cavity monitors, to tune the op-amps to achieve a -50 mV signal from each of the cavity monitor signals. The table below shows the voltage output from the op-amp modules, which then goes into the ADC in ROC2 Slot 23 channels 32-37:

    Cavity Signal           Output Voltage (mV)
    -------------------------------------------
    Upstream Intensiy           -51.6
    Upstream X pos              -49.6
    UPstream Y pos              -49.5
    Downstream Intensity        -49.9
    Downstream X pos            -49.7
    Downstream Y pos            -49.4
    

    2) Last night and this morning I tested the response of the op-amps and ADC channels. The expected response was observed. With less attenuation, i.e. a larger negative signal, the signal in the ADC appears at larger ADC channel, and with more attenuation, the signal appears at lower ADC channel.

    3) While I was testing the downstream op-amp module this morning, I saw that the upstream ADC channels were saturating at low channel. I checked the signal with a DVM and found that the op-amps for the upstream cavity were producing a +10.4 V output from all four channels. I reset the crate and unplugged the inputs to this module, but the +10.4 V output remained. The other four modules in the same crate are working fine. Hence, I believe this module is broken.

    I was unable to find a spare op-amp module, so I used three unused channels in the four remaining boards: 1) one from the downstream cavity monitor module 2) two from the raster module

    Afterwards, I fine-tuned and checked the signals for all six cavity monitor channels. The settings still need to be tested with beam to ensure that everything is working fine.

    A copy of this log entry has been emailed to: rom@jlab.org,bdhahn@jlab.org