User name R. Michaels
Log entry time 11:51:58 on April11,2008
Entry number 231382
Followups:
keyword=access: MPS, scaler gate, T7
We flipped the MPS polarity. Actually it had been flipped
accidentally by someone on Apr 3 -- the fiber converter has 4
outputs, 2 of one polarity and 2 of the other (and not labelled !).
I took access and put the scaler gate back to normal polarity.
I also fixed T7. Actually it has been wrong since the start, I think.
Although I had checked it, somehow one of the ECL cables disappeared
so T7 involved only the 4th paddle of S1.
Now T7 involves paddles 2 through 5 of S1.
I verified by turning HV on/off on individual channels that T7 does
not involve paddles 1 or 6, but is the "or" of 2-5.
It should be checked with simple offline analysis (I guess this
was not done before ??).