Main INDEX
Monthly INDEX
PREV
User name V. Sulkosky
Log entry time 18:45:53 on July 31, 2008
Entry number 241970
keyword=Pulser Test of S2m Electronics and Cabling
Today we tested the electronics including the discriminators using
pocket pulsers and a PS amplifier plugged into the EDTM cables that
connect to the splitter box. All channels from the discriminator
had a signal. The discriminators' threshold value was checked and
measured to be -9.8 mV for both modules.
The next test is to turn the HV on for S1 and S2m to check the signals
from the PMTs, the timing between S1 and S2m and the rate from cosmics.