T2, T6, and T7 should all count the same, since the HV on BigBite is turned off and they all are triggering on edtm pulser. From the summary below, it is indeed the case for the "bbite" scalers but not for the "left".
Note : On left arm, for T7, (+-) and (--) counts match with bbite whereas (++) counts twice and (-+) is completely different from bbite.
I will check them tomorrow when the hall opens and see whats is causing this double counts on these triggers.
------------- Scaler Summary ---------------- Scaler bank bbite clock location 4 7 clock counts 491050 clock rate 1024 Time of run 7.99 min Triggers: 1 = 5995 2 = 5995 3 = 1071927 4 = 30909 5 = 5989 Accepted triggers: 168600 Accepted triggers by target and helicity state: (++) = 35239 (+-) = 34884 (-+) = 42042 (--) = 41923 Trigger 1,2,3,4,5,6,7,8 by target and helicity state : (++) (+-) (-+) (--) trig 1 1341 1375 1291 1342 trig 2 1341 1375 1291 1342 trig 3 232989 233047 245625 245668 trig 4 6711 6741 7091 7032 trig 5 1339 1374 1291 1341 trig 6 1341 1375 1291 1342 trig 7 1341 1375 1291 1342 trig 8 110929 110905 107869 107975 Charge Monitors (Micro Coulombs) Upstream BCM gain x1 3392.96 x3 3340.45 x10 -1.42 Downstream BCM gain x1 3496.13 x3 3477.34 x10 3477.34 [adaq@adaql4 scaler]$ ./tscalntup /adaql2/data3/e06010_2180.dat.0 Left Bank = Left Opened scaler map file scaler.mapBdate : day = 26 month = 8 year = 2008 Using Left Arm BCM pedestals Normal end of file /adaql2/data3/e06010_2180.dat.0 encountered
------------- Scaler Summary ---------------- Scaler bank Left clock location 1 7 clock counts 480074 clock rate 1024 Time of run 7.81 min Triggers: 1 = 5861 2 = 5861 3 = 1056147 4 = 30174 5 = 5856 Accepted triggers: 168853 Accepted triggers by target and helicity state: (++) = 35297 (+-) = 34933 (-+) = 42104 (--) = 41987 Trigger 1,2,3,4,5,6,7,8 by target and helicity state : (++) (+-) (-+) (--) trig 1 1341 1375 1292 1342 trig 2 2682 2748 2582 2683 trig 3 235325 235454 248110 248067 trig 4 6715 6741 7098 7031 trig 5 1339 1374 1291 1341 trig 6 2682 2749 2582 2683 trig 7 2683 1375 2079 1342 trig 8 110928 110905 107867 107961 Charge Monitors (Micro Coulombs) Upstream BCM gain x1 3308.58 x3 3257.20 x10 -1.38 Downstream BCM gain x1 3408.95 x3 3390.56 x10 3390.55
A copy of this log entry has been emailed to: rom@jlab.org