    User name brads

    Log entry time 01:37:27 on October 30, 2008

    Entry number 245466

    This entry is a followup to: 245459

    Followups:

    keyword=BERR light illuminated on 1151 scaler in ROC9

    Sigh... The 'BERR' (Branch error) light has lit up on the 1151 scaler in ROC9. We tried reseating the scaler and power cycling the crate but it won't turn off.

    Kalyan checked the scaler data and they seem to be good (no analyzer errors, no errors on the VME console, and the sync scaler data have a 1:1 relationship with that of three other ROCs. We can't find a spare 1151 and didn't have another SFI handy so we will leave it the way it is for tonight.

    I found a second SFI in the cabinet upstairs and will track down another 1151 tomorrow morning. We'll do a swap at the first convenient access tomorrow. I made a note on the whiteboard so we don't forget.

    A copy of this log entry has been emailed to: brads, moffit