    User name Jack

    Log entry time 12:08:55 on January 12, 2009

    Entry number 255551

    keyword=Right Arm VDC bad HV cable

    One of the SHV connectors on the high voltage cable from the power supply to the chamber was bad. The VDC had no high voltage on the wires, hence no output. I replaced the connector. It looked like the old connector had the dielectric too long and the braid was barely crimped in the collar/ferrule. Over time it must have worked loose. The connector up by the Bertan supply, labeled "Bottom", also looks questionable and needs to be re-made at some point.