• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name Chiranjib

    Log entry time 19:38:52 on January 12, 2009

    Entry number 255659

    keyword=Internal Temperature Test ( with Al's suggestions)

    Another internal temp. test is performed after discussing with Al. This method has been applied at UVA. Since we have some time, we could afford to loose some polarization.

    Cell :: BRADY Holding field :: Transverse - Pumping :: - NMR frequency sweep done on pumping chamber. RF = longitudinal pick up (PC) = vertical

    *************************************************************** Data :: C:\HappyValentine\NMR\Data\NMR_Freq_He_20090112_191755 ***************************************************************

    15 NMR frequency sweeps have been done in the following order---

    LASER ON ------> NMR1 , NMR2 , NMR3 LASER OFF------> NMR4 , NMR5 , NMR6 LASER ON ------> NMR7 , NMR8 , NMR9 LASER OFF------> NMR10 , NMR11 , NMR12 LASER ON ------> NMR13 , NMR14 , NMR15

    The time interval between each NMR is 1 min.



    A copy of this log entry has been emailed to: jpchen,yqiang