• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name Chiranjib

    Log entry time 19:38:58 on January 12, 2009

    Entry number 255660

    keyword=Internal Temperature Test ( with Al's suggestions)

    Another internal temp. test is performed after discussing with Al. This
    method has been applied at UVA. Since we have some time, we could afford
    to loose some polarization.

    Cell :: BRADY
    Holding field :: Transverse -
    Pumping :: -
    NMR frequency sweep done on pumping chamber.
    RF = longitudinal
    pick up (PC) = vertical

    ***************************************************************
    Data :: C:\HappyValentine\NMR\Data\NMR_Freq_He_20090112_191755
    ***************************************************************

    15 NMR frequency sweeps have been done in the following order---

    LASER ON ------> NMR1 , NMR2 , NMR3
    LASER OFF------> NMR4 , NMR5 , NMR6
    LASER ON ------> NMR7 , NMR8 , NMR9
    LASER OFF------> NMR10 , NMR11 , NMR12
    LASER ON ------> NMR13 , NMR14 , NMR15

    The time interval between each NMR is 1 min.





    A copy of this log entry has been emailed to: jpchen,yqiang