Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name Kalyan
Log entry time 19:13:02 on January16,2009
Entry number 256375
keyword=L-HRS EDTM signal on TDC have some problem
Starting from run 5194 the Left arm EDTM signal in the TDC is having some problem. Even after having the no EDTM cut on the coin. time we are seeing some EDTM events in the spectrum(shown below run 5195, you can see some pulser events around chan 975).
Earlier runs had no such problem when we apply no EDTM cut.
The BigBite copy of the EDTM in the TDC is still fine and works as expected when used in the cuts. For now we can use this copy to cut out EDTM.
FIGURE 1