    keyword=Beam Position scan to pin down He3 cell center

    A BPM position scan is performed between 4:43Am and 5:19AM. The purpose of the test is to find the center position of the cell relative to EPICS read back of BPM. The conclusions are


    Raw Data

    The scan is performed with max raster 5x4mm (6x6 MCC) by

    1. Spot++ confirming not hitting cell wall. Mark norminal trigger rate
    2. Scan on BPMX untill reach both cell wall (tell by trigger rate). Keeping BPMA_X-BPMB_X = -0.5mm (might not be optimal difference)
    3. Center X, scan BPMYs, Keeping BPMA_Y = BPMB_Y (might not be optimal difference)
    4. Center Y
    Scan ID BPM A X BPMA Y BPMB X BPMB Y T1/10000 T3/100 Average X (mm) Average Y (mm) T1*T3(1M/s^2)
    1 -1.022 2.175 -0.405 2.216 2.324 2.95 -0.7135 2.1955 6.8558
    3 -0.5 2.175 0 2.216 2.3 2.9 -0.25 2.1955 6.67
    4 0 2.175 0.5 2.216 2.4 2.9 0.25 2.1955 6.96
    5 0.3 2.175 0.8 2.216 2.5 3.5 0.55 2.1955 8.75
    6 0 2.175 -0.6 2.216 2.3 3 -0.3 2.1955 6.9
    7 -1.8 2.175 -1.38 2.216 2.3 3 -1.59 2.1955 6.9
    8 -3 2.175 -2.4 2.216 2.6 3.7 -2.7 2.1955 9.62
    9 -1.3 2.175 -0.8 2.216 2.2 2.8 -1.05 2.1955 6.16
    10 -1.3 1 -0.8 1 2.3 3.1 -1.05 1 7.13
    11 -1.3 0 -0.8 0 2.5 3.4 -1.05 0 8.5
    12 -1.3 3 -0.8 3 2.3 2.9 -1.05 3 6.67
    13 -1.3 4 -0.8 4 2.3 2.9 -1.05 4 6.67
    14 -1.3 5 -0.8 5 2.5 3.4 -1.05 5 8.5
    15 -1.3 2.5 -0.8 2.5 2.3 2.9 -1.05 2.5 6.67
    2 0.62 2.175 1.6 2.216 17 133 1.11 2.1955 2261

    Scan Plots + Fitting



    A copy of this log entry has been emailed to: jpchen



    trigger rate vs BPMX



    trigger rate vs BPMY